FDA Adverse Event Injury Summary report: N

TUBING SET

MDR report key: 24145549 · Received January 22, 2026

Report

Report Number
3013876692-2026-00007
Event Type
Injury
Date Received
January 22, 2026
Date of Event
September 1, 2025
Report Date
February 26, 2026
Manufacturer
MAQUET CARDIOPULMONARY GMBH
Product Code
DTZ
UDI-DI
04058863005744
PMA / PMN Number
K112360
Adverse Event
Yes
Report Source
Distributor report
Reporter Location
US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

A FOLLOW-UP MEDWATCH WILL BE SUBMITTED WHEN ADDITIONAL INFORMATION BECOMES AVAILABLE. NOTE: THIS EVENT OCCURRED ON THE SPANISH MARKET. IT IS A SIGNIFICANTLY SIMILAR DEVICE TO "HLS SET¿ WHICH IS SOLD IN THE USA UNDER PREMARKET SUBMISSION NUMBER K112360. FOR SECTION D4 ALL AVAILABLE IDENTIFYING INFORMATION IS FOR THE OUT OF THE US DEVICE, SUBJECTED TO THIS REPORT. THEREFORE, NO GUDID INFORMATION EXISTS. FURTHERMORE, UDI INFORMATION (PRIMARY DI NUMBER) PROVIDED IN D4 IS FOR HLS SET WITH CATALOG NUMBER 701069073.

Additional Manufacturer Narrative · 0

MAQUET CARDIOPULMONARY GMBH HAS RECEIVED THE FOLLOWING INFORMATION FROM THE SPANISH AGENCY OF MEDICINES AND MEDICAL DEVICES ON 2026-01-19: THE VIGILANCE AREA OF MEDICAL DEVICES OF THE SPANISH AGENCY OF MEDICINES AND MEDICAL DEVICES HAS BEEN INFORMED THROUGH RENACER REGISTRY ABOUT AN INCIDENT INVOLVING THE PRODUCT ¿CARDIOHELP¿, MANUFACTURED BY GETINGE AND DISTRIBUTED BY, OCCURRED ON 2025-09-01. THE INFORMATION WAS PROVIDED THAT THE PATIENT HAD THE INCIDENT "SUSPECTED COMPARTMENT SYNDROME IN THE RIGHT UPPER LIMB EXTENDING TO THE NECK AND IPSILATERAL PECTORAL MUSCLE. EXTRAVASATION OF REPERFUSION CANNULA ". NEW INFORMATION WAS RECEIVED ON 2026-02-06 THAT THE PATIENT EXPERIENCED DURING ECMO SUPPORT A SUSPECTED COMPLICATION OF THE RIGHT UPPER EXTREMITY COMPARTMENT SYNDROM (RETURN CANNULA IN THE RIGHT SUBCLAVIAN ARTERY) STATED BY THE CUSTOMER. THE SURGICAL EXAMINATION REVEALED A REPERFUSION CANNULA WITH ACTIVE BACKFLOW BLEEDING. IT WAS DECIDED BY THE CUSTOMER TO REMOVE IT AND CLOSE IT WITH SUTURES. THE CUSTOMER CONFIRMED THAT THERE WAS NO FAILURE OF THE GETINGE DEVICES USED. FOLLOWING PATIENT DATA WAS SHARED: 47 YEAR OLD MALE WITH 97KG. NO DEVICE HISTORY REVIEW, NO REVIEW OF SCRAP, REWORK, ENHANCEMENTS AND DESIGN CHANGES WAS PERFORMED, AS NO FAILURE OF THE DEVICE WAS REPORTED. IN ADDITION, NO REVIEW FOR POTENTIAL FIELD ACTIONS AND CAPAS RELATED TO THE EVENT AND PRODUCT IN THIS COMPLAINT AND NO REVIEW OF THE NON-CONFORMITIES HAS BEEN PERFORMED, AS THIS COMPLAINT WAS REPORTED DUE TO THE EVENT AND NOT DUE TO A FAILURE ON A GETINGE PRODUCT. BASED ON THE AVAILABLE INFORMATION THE REPORTED EVENT COULD NOT BE LINKED TO A DEVICE DEFECT/FAILURE. THEREFORE THE COMPLAINT COULD NOT BE CONFIRMED. THE OCCURRENCE RATE RELATED TO THE REPORTED ISSUE IS CURRENTLY BEING MONITORED AS PART OF MAQUET CARDIOPULMONARY¿ S TRENDING PROGRAM AND ADDITIONAL INVESTIGATIONS OR CORRECTIONS WILL BE IMPLEMENTED IN CASE OF ADVERSE TRENDING.

Description of Event or Problem · 0

MAQUET CARDIOPULMONARY GMBH HAS RECEIVED THE FOLLOWING INFORMATION FROM THE SPANISH AGENCY OF MEDICINES AND MEDICAL DEVICES ON (B)(6) 2026: THE VIGILANCE AREA OF MEDICAL DEVICES OF THE SPANISH AGENCY OF MEDICINES AND MEDICAL DEVICES HAS BEEN INFORMED THROUGH RENACER REGISTRY ABOUT AN INCIDENT INVOLVING THE PRODUCT ¿CARDIOHELP¿, MANUFACTURED BY GETINGE AND DISTRIBUTED BY, OCCURRED ON (B)(6) 2025 WITH REPORT NUMBER (B)(4). THE INFORMATION WAS PROVIDED THAT THE PATIENT HAD THE INCIDENT "SUSPECTED COMPARTMENT SYNDROME IN THE RIGHT UPPER LIMB EXTENDING TO THE NECK AND IPSILATERAL PECTORAL MUSCLE. EXTRAVASATION OF REPERFUSION CANNULA ". AS THE PATIENT SUFFERED A SERIOUS INJURY, A REPORT IS REQUIRED. COMPLAINT ID# (B)(4).

Description of Event or Problem · 0

COMPLAINT ID (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
210362 TUBING SET OXYGENATOR, CARDIOPULMONARY BYPASS DTZ MAQUET CARDIOPULMONARY GMBH HLS SET 3000459825 04058863005744

Patients

Seq Age Sex Outcome Treatment
1 47 YR Male Life Threatening| O