FDA Adverse Event Malfunction Summary report: N

ENDOWRIST SP

MDR report key: 24142029 · Received January 22, 2026

Report

Report Number
2955842-2026-01849
Event Type
Malfunction
Date Received
January 22, 2026
Date of Event
December 29, 2025
Report Date
January 21, 2026
Manufacturer
INTUITIVE SURGICAL, INC
Product Code
NAY
UDI-DI
00886874114285
PMA / PMN Number
K182371
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

INTUITIVE SURGICAL, INC. (ISI) DID RECEIVE THE SINGLE PORT MONOPOLAR CURVED SCISSORS (MCS) INSTRUMENT TO PERFORM FAILURE ANALYSIS. THE INSTRUMENT WAS FOUND TO HAVE A BROKEN INSTRUMENT TUBE ADAPTER. THIS COMPONENT IS LOCATED AT THE DISTAL END OF THE INSTRUMENT AND SERVES AS THE INTERFACE BETWEEN THE INSTRUMENT AND THE USER INSTALLED TIP ACCESSORY. THE BROKEN PIECE WAS NOT RETURNED AND WOULD HAVE MEASURED APPROXIMATELY 2.08MM X 1.74MM. THE BROKEN TIP ADAPTER MADE IT DIFFICULT TO INSTALL A TIP AND CAUSED THE TIP TO NOT BE SECURELY ATTACHED. THE PROBABLE ROOT CAUSE OF A DAMAGED TUBE ADAPTER IS ATTRIBUTED TO EITHER TIP ACCESSORY INSTALLATION ISSUES OR DAMAGE DURING USE.

Description of Event or Problem · 0

IT WAS REPORTED THAT PRIOR TO A DA VINCI-ASSISTED SURGICAL PROCEDURE, THE TIP OF THE SINGLE PORT MONOPOLAR CURVED SCISSORS INSTRUMENT COULD NOT BE LOADED. THE DISTAL END WAS CRACKED. THE PROCEDURE WAS COMPLETED WITH NO REPORTED INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
203657 ENDOWRIST SP MONOPOLAR CURVED SCISSORS NAY INTUITIVE SURGICAL, INC 430004-57 U10230720 0009 00886874114285

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown DA VINCI INSTRUMENTS AND ACCESSORIES