STYLE 15 SILICONE GEL FILLED BREAST IMPLANT
Report
- Report Number
- 9617229-2026-01301
- Event Type
- Injury
- Date Received
- January 21, 2026
- Date of Event
- December 16, 2025
- Report Date
- April 23, 2026
- Manufacturer
- ALLERGAN (COSTA RICA)
- Product Code
- FTR
- UDI-DI
- 10888628001480
- PMA / PMN Number
- P020056
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WI, US
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
A REVIEW OF THE DEVICE HISTORY RECORD HAS BEEN COMPLETED. NO DEVIATIONS OR NON-CONFORMANCES NOTED. FURTHER INFORMATION FROM THE REPORTER REGARDING EVENT, PRODUCT, OR PATIENT DETAILS HAS BEEN REQUESTED. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME. REASON FOR REOPERATION: RUPTURE.
DEVICE EVALUATION: THE DEVICE RELATED TO THE REPORTED EVENT OF RUPTURE WAS RECEIVED ON MARCH 27, 2026, WITH LOT NUMBER 1810111. PER THE INVESTIGATION PROCEDURE, THE DEVICE IS ANALYZED THROUGH VISUAL INSPECTION MICROSCOPIC INSPECTION IF OPENINGS ARE OBSERVED AND A WEIGHT VERIFICATION. PER THE ANALYSIS PERFORMED, THE ASSESSMENTS OF THE COMPLAINTS AND ANY POTENTIAL MANUFACTURING ISSUE ARE DISPLAYED ALONG WITH ANY FURTHER ACTIONS REQUIRED: ¿ RUPTURE: OBSERVED BROKEN DEVICE THROUGH MICROSCOPIC INSPECTION ASSESSED AS UNIDENTIFIED (TEAR) OPENING AND MISSING PIECE OF SHELL ASSESSED AS INCONCLUSIVE. NO FURTHER ACTIONS ARE REQUIRED AS IT IS NOT RELATED TO THE MANUFACTURING PROCESS. NONE OF THE OTHER OBSERVATIONS PERFORMED DURING THE DEVICE ANALYSIS (CREASES) ARE FOUND TO BE POTENTIALLY RELATED TO THE MANUFACTURING PROCESS, AND, THEREFORE, NO FURTHER ACTIONS ARE REQUIRED FOR THESE OBSERVATIONS.
HEALTHCARE PROFESSIONAL REPORTED "RUPTURE". THIS RECORD IS FOR THE RIGHT SIDE. THE DEVICE WAS EXPLANTED.
HEALTHCARE PROFESSIONAL REPORTED "RUPTURE". THIS RECORD IS FOR THE RIGHT SIDE. THE DEVICE WAS EXPLANTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 198527 | STYLE 15 SILICONE GEL FILLED BREAST IMPLANT | PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED | FTR | ALLERGAN (COSTA RICA) | 1810111 | 10888628001480 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 63 YR | Female | Required Intervention |