FDA Adverse Event Injury Summary report: N

STYLE 15 SILICONE GEL FILLED BREAST IMPLANT

MDR report key: 24137149 · Received January 21, 2026

Report

Report Number
9617229-2026-01301
Event Type
Injury
Date Received
January 21, 2026
Date of Event
December 16, 2025
Report Date
April 23, 2026
Manufacturer
ALLERGAN (COSTA RICA)
Product Code
FTR
UDI-DI
10888628001480
PMA / PMN Number
P020056
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WI, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

A REVIEW OF THE DEVICE HISTORY RECORD HAS BEEN COMPLETED. NO DEVIATIONS OR NON-CONFORMANCES NOTED. FURTHER INFORMATION FROM THE REPORTER REGARDING EVENT, PRODUCT, OR PATIENT DETAILS HAS BEEN REQUESTED. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME. REASON FOR REOPERATION: RUPTURE.

Additional Manufacturer Narrative · 0

DEVICE EVALUATION: THE DEVICE RELATED TO THE REPORTED EVENT OF RUPTURE WAS RECEIVED ON MARCH 27, 2026, WITH LOT NUMBER 1810111. PER THE INVESTIGATION PROCEDURE, THE DEVICE IS ANALYZED THROUGH VISUAL INSPECTION MICROSCOPIC INSPECTION IF OPENINGS ARE OBSERVED AND A WEIGHT VERIFICATION. PER THE ANALYSIS PERFORMED, THE ASSESSMENTS OF THE COMPLAINTS AND ANY POTENTIAL MANUFACTURING ISSUE ARE DISPLAYED ALONG WITH ANY FURTHER ACTIONS REQUIRED: ¿ RUPTURE: OBSERVED BROKEN DEVICE THROUGH MICROSCOPIC INSPECTION ASSESSED AS UNIDENTIFIED (TEAR) OPENING AND MISSING PIECE OF SHELL ASSESSED AS INCONCLUSIVE. NO FURTHER ACTIONS ARE REQUIRED AS IT IS NOT RELATED TO THE MANUFACTURING PROCESS. NONE OF THE OTHER OBSERVATIONS PERFORMED DURING THE DEVICE ANALYSIS (CREASES) ARE FOUND TO BE POTENTIALLY RELATED TO THE MANUFACTURING PROCESS, AND, THEREFORE, NO FURTHER ACTIONS ARE REQUIRED FOR THESE OBSERVATIONS.

Description of Event or Problem · 0

HEALTHCARE PROFESSIONAL REPORTED "RUPTURE". THIS RECORD IS FOR THE RIGHT SIDE. THE DEVICE WAS EXPLANTED.

Description of Event or Problem · 0

HEALTHCARE PROFESSIONAL REPORTED "RUPTURE". THIS RECORD IS FOR THE RIGHT SIDE. THE DEVICE WAS EXPLANTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
198527 STYLE 15 SILICONE GEL FILLED BREAST IMPLANT PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED FTR ALLERGAN (COSTA RICA) 1810111 10888628001480

Patients

Seq Age Sex Outcome Treatment
1 63 YR Female Required Intervention