WAVEWRITER ALPHA?
Report
- Report Number
- 3006630150-2026-00442
- Event Type
- Injury
- Date Received
- January 21, 2026
- Date of Event
- December 30, 2025
- Report Date
- April 2, 2026
- Manufacturer
- BOSTON SCIENTIFIC NEUROMODULATION CORPORATION
- Product Code
- LGW
- UDI-DI
- 08714729985099
- PMA / PMN Number
- P030017
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AR, US
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
MODEL NUMBER/CATALOG NUMBER: SC-1232/ SC-2317-50/ SC-4318. SERIAL NUMBER: (B)(6). BATCH/LOT NUMBER: 767898/ 7082103/ 7082174/ 34479890. INVESTIGATION RESULTS. THE DEVICE WAS NOT RETURNED FOR ANALYSIS; THEREFORE, A TECHNICAL ANALYSIS COULD NOT BE PERFORMED. GOOD FAITH EFFORT ATTEMPTS WERE MADE TO TRY AND RETRIEVE THE DEVICE, HOWEVER RESPONSE WAS NOT RECEIVED. DEVICE HISTORY RECORD . IT WAS CONFIRMED THIS DEVICE MET MANUFACTURING SPECIFICATION PRIOR TO DISTRIBUTION AND THERE WERE NO MANUFACTURING DEVIATIONS WHICH COULD HAVE CONTRIBUTED TO THE REPORTED EVENT. LABELING REVIEW . REVIEW OF THE INSTRUCTIONS FOR USE (IFU) CONFIRMED THERE WAS RELEVANT CONTENT AND SUFFICIENT GUIDANCE WITH RESPECT TO THE CIRCUMSTANCES DESCRIBED WITHIN THIS COMPLAINT. NO UPDATES ARE REQUIRED TO THE IFU AS A RESULT OF THIS EVENT. INVESTIGATION CONCLUSION. BASED ON A THOROUGH REVIEW OF THE REPORTED COMPLAINT, BOSTON SCIENTIFIC HAS ASSIGNED AN INVESTIGATION CONCLUSION CODE OF KNOWN INHERENT RISK OF DEVICE.
BLOCK B3: APPROXIMATED BASED ON THE DATE THE MANUFACTURER BECAME AWARE OF THE EVENT. ADDITIONAL SUSPECT MEDICAL DEVICE COMPONENT INVOLVED IN THE EVENT: PRODUCT FAMILY: SCS-LINEAR LEADS. UPN: M365SC2317500. MODEL: SC-2317-50. SERIAL: (B)(6). BATCH: (B)(6). UDI: (B)(4). PRODUCT FAMILY: SCS-LINEAR LEADS. UPN: M365SC2317500. MODEL: SC-2317-50. SERIAL: (B)(6). BATCH: (B)(6). UDI: (B)(4). PRODUCT FAMILY: SCS-LINEAR LEADS. UPN: M365SC43180. MODEL: SC-4318. BATCH: 34479890. UDI: (B)(4).
IT WAS REPORTED THAT THE PATIENT EXPERIENCED PAIN AT THE POCKET SITE DUE TO AN INFECTION THAT HAD DEVELOPED AT THE AREA. THE PATIENT UNDERWENT A SPINAL CORD STIMULATOR (SCS) EXPLANT PROCEDURE. THE PHYSICIAN IRRIGATED AND CLOSED THE WOUND SITE. THE PATIENT WAS PRESCRIBED WITH ANTIBIOTICS AND WAS DOING WELL POSTOPERATIVELY. THE PHYSICIAN CONFIRMED THAT THE INFECTION WAS NEITHER DEVICE NOR PROCEDURE RELATED. THE EXPLANTED DEVICES WERE KEPT BY THE FACILITY.
IT WAS REPORTED THAT THE PATIENT EXPERIENCED PAIN AT THE POCKET SITE DUE TO AN INFECTION THAT HAD DEVELOPED AT THE AREA. THE PATIENT UNDERWENT A SPINAL CORD STIMULATOR (SCS) EXPLANT PROCEDURE. THE PHYSICIAN IRRIGATED AND CLOSED THE WOUND SITE. THE PATIENT WAS PRESCRIBED WITH ANTIBIOTICS AND WAS DOING WELL POSTOPERATIVELY. THE PHYSICIAN CONFIRMED THAT THE INFECTION WAS NEITHER DEVICE NOR PROCEDURE RELATED. THE EXPLANTED DEVICES WERE KEPT BY THE FACILITY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 17291 | WAVEWRITER ALPHA? | STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF | LGW | BOSTON SCIENTIFIC NEUROMODULATION CORPORATION | SC-1232 | 767898 | 08714729985099 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 61 YR | Female | Required Intervention |