FDA Adverse Event
Other
Summary report: N
FMP HIP
MDR report key: 2413059
·
Received January 6, 2012
Report
- Report Number
- 1644408-2011-00698
- Event Type
- Other
- Date Received
- January 6, 2012
- Date of Event
- December 21, 2011
- Report Date
- December 21, 2011
- Manufacturer
- ENCORE MEDICAL, L.P.
- Product Code
- KWA
- PMA / PMN Number
- K003250
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CO, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Description of Event or Problem · 1
REVISION SURGERY - THE PT HAD RECURRENT DISLOCATION FROM IMPINGEMENT OF THE FEMORAL NECK ON POSTERIOR/INFERIOR RIM OF THE ACETABULAR LINER. THE SURGEON REPORTED NOTCHING OF THE FEMORAL STEM DUE TO METAL TRANSFER, AS WELL AS THE METAL INSERT "SPINNING" IN THE POLYETHYLENE. THE SURGEON REVISED TO A LARGER CERAMIC HEAD AND POLY LINER.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | FMP HIP | COCR HEAD NEUTRAL | KWA | ENCORE MEDICAL, L.P. | 692391 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 58 YR | 499-28-006, LOT 596661 |