Narratives
Description of Event or Problem · 0
A VENTILATOR WAS RETURNED TO THE MANUFACTURER FOR SERVICE. THERE WAS NO HARM OR INJURY REPORTED. DURING THE EVALUATION OF THE DEVICE AT THE MANUFACTURER'S SERVICE CENTER, THE DEVICE FAILED A TEST STEP FOR OXYGEN LOW FLOW. THE DEVICE'S INLET AIR PATH, AIR PATH FOAM, AND FLOW PATH WERE REPLACED TO ADDRESS THE ISSUE. THE DEVICE PASSED ALL TESTING.