DENTAL SCREW
Report
- Report Number
- 0001038806-2026-00374
- Event Type
- Malfunction
- Date Received
- January 21, 2026
- Report Date
- January 21, 2026
- Manufacturer
- ZIMVIE US CORP LLC
- Product Code
- DZE
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
ZIMVIE COMPLAINT NUMBER (B)(4). ZIMVIE RECEIVED AN UNKNOWN ZIMMER SCREW FOR EVALUATION. VISUAL EVALUATION OF THE AS RETURNED DEVICES IDENTIFIED THE FRACTURED SCREW FRAGMENT STUCK INSIDE THE IMPLANT. FUNCTIONAL TESTING TO RECREATE THE REPORTED EVENT COULD NOT BE PERFORMED DUE TO THE NATURE OF THE DEVICE & EVENT. DHR, STERILIZATION, AND COMPLAINT HISTORY REVIEW COULD NOT BE PERFORMED, AS THE SUBJECT LOT NUMBER ASSOCIATED WITH THE UNKNOWN ZIMMER SCREW IS NOT AVAILABLE. ZIMVIE QUALITY MANAGEMENT SYSTEM (QMS) HAS CONTROLS IN PLACE TO PREVENT THE DISTRIBUTION OF NON-CONFORMING PRODUCT AND ENSURE THE PRODUCT IS WITHIN SPECIFICATIONS. BASED ON THE INVESTIGATION AND RISK MANAGEMENT FILE REVIEW, THE MOST LIKELY ROOT CAUSES DETERMINED FROM THE INVESTIGATION ARE CLINICIAN ERROR, OR PATIENT FACTORS. THEREFORE, BASED ON THE AVAILABLE INFORMATION, DEVICE MALFUNCTION DID OCCUR. THE REPORTED EVENT WAS CONFIRMED WITH ALL THE AVAILABLE INFORMATION. NO FURTHER INVESTIGATION OR IMMEDIATE CAPA / HHE/D ESCALATION IS REQUIRED, AS THE COMPLAINT INVESTIGATION DID NOT CONFIRM THE PRODUCT WAS NONCONFORMING AT THE TIME OF DISTRIBUTION, AND NO NEW FAILURE MODE, HARM, OR HAZARDOUS SITUATION WAS IDENTIFIED THROUGH THE INVESTIGATION PERFORMED.
DURING INVESTIGATION, A FRACTURED SCREW WAS FOUND INSIDE OF THE IMPLANT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 192641 | DENTAL SCREW | DZE | ZIMVIE US CORP LLC |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 83 YR | Male | TSVB11, IMPL TAPERED SCR-V MTX 3.7MM 3.5MM 11.5MM |