I-STAT TROPONIN (HS-TNI ) CARTRIDGE

MDR report key: 24119055 · Received January 20, 2026

Report

Report Number
2245578-2026-00017
Event Type
Malfunction
Date Received
January 20, 2026
Date of Event
January 12, 2026
Report Date
March 12, 2026
Manufacturer
ABBOTT POINT OF CARE INC.
Product Code
MMI
PMA / PMN Number
K240984
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

APOC INCIDENT # (B)(4). APOC LABELING WILL BE EVALUATED DURING THE INVESTIGATION AS PERTAINING TO THE EVENT.

Additional Manufacturer Narrative · 0

APOC INCIDENT: (B)(4). THE INVESTIGATION WAS COMPLETED ON 05-MAR-2026. A REVIEW OF THE DEVICE HISTORY RECORD CONFIRMED THE CARTRIDGE LOT MET FINISHED GOODS RELEASE CRITERIA. RETAINED CARTRIDGE TESTING MET THE ACCEPTANCE CRITERIA OUTLINED IN APPENDIX 1 OF Q04.01.003 REV. AP (PRODUCT COMPLAINT LEVEL 2 AND LEVEL 3 INVESTIGATION PROCEDURE). NO DEFICIENCY HAS BEEN IDENTIFIED FOR HS-TNI CARTRIDGE LOT B25319.

Description of Event or Problem · 0

ON 15-JAN-2025, ABBOTT POINT OF CARE WAS CONTACTED BY A CUSTOMER REGARDING I-STAT TROPONIN CARTRIDGES THAT YIELDED DISCREPANT FALSE POSITIVE RESULTS ON A 22 YEAR FEMALE PATIENT BACK & CHEST PAIN. THERE WAS NO ADDITIONAL INFORMATION. METHOD DATE, TESTED RESULTS (NG/L), SAMPLE POSITIVE/NEGATIVE: I-STAT (B)(6) 2026 , 19:30 >1000, A POSITIVE. I-STAT (B)(6) 2026, 19:55 >1000, B POSITIVE. POSITIVE CUT-OFF VALUE FOR I-STAT TROPONIN TEST: 13. FINAL PATIENT DIAGNOSIS: PRECORDIAL PAIN. THERE ARE NO INJURIES ASSOCIATED WITH THIS EVENT. AT THIS TIME THERE IS NO REASON TO SUSPECT A MALFUNCTION EXISTS. THE INVESTIGATION IS UNDERWAY. 99TH PERCENTILE 90% CI. SEX: N. (NG/L, PG/ML). (NG/L, PG/ML). FEMALE: 490, 13 (10, 17). MALE: 404, 28 (19, 58). OVERALL: 895, 21 (14, 30).

Description of Event or Problem · 0

NA.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
55244 I-STAT TROPONIN (HS-TNI ) CARTRIDGE HS-TNI CARTRIDGE MMI ABBOTT POINT OF CARE INC. NA B25319

Patients

Seq Age Sex Outcome Treatment
1 22 YR Female