FDA Adverse Event Malfunction Summary report: N

ACCESS

MDR report key: 2411692 · Received January 11, 2012

Report

Report Number
6000001-2012-02234
Event Type
Malfunction
Date Received
January 11, 2012
Date of Event
December 19, 2011
Report Date
December 19, 2011
Manufacturer
BAXTER HEALTHCARE - MALTA
Product Code
LJH
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE SAMPLE IS REPORTED TO BE AVAILABLE BUT HAS NOT BEEN RECEIVED FOR EVALUATION. IF ADDITIONAL INFORMATION BECOMES AVAILABLE OR THE SAMPLE IS RECEIVED, A FOLLOW UP REPORT WILL BE SUBMITTED. BAXTER WILL CONTINUE TO MONITOR SIMILAR REPORTS TO DETERMINE IF FURTHER ACTIONS ARE REQUIRED. THIS DEVICE IS MANUFACTURED FOR DISTRIBUTION OUTSIDE OF THE UNITED STATES (US); THEREFORE, IT DOES NOT CONTAIN A US 510K NUMBER. HOWEVER, THIS MDR IS BEING SUBMITTED BECAUSE IT IS THE SAME AS OR SIMILAR TO A PRODUCT DISTRIBUTED WITHIN THE US.

Additional Manufacturer Narrative · 1

(B)(4). THREE COMPANION SAMPLES WERE RETURNED TO THE MANUFACTURING PLANT FOR EVALUATION. VISUAL INSPECTION WAS PERFORMED AND NO DEFECTS WERE OBSERVED. THE SETS WERE TESTED FOR GRAVITY AND THEN FOR LEAKAGE AT 0.56 BAR; NO ISSUE WAS OBSERVED. THEREFORE, THE REPORTED CONDITION WAS NOT CONFIRMED. AN ASSIGNABLE ROOT CAUSE COULD NOT BE DETERMINED. A BATCH REVIEW WAS PERFORMED ON THE REPORTED LOT WITH NO DEVIATIONS FOUND. THIS ISSUE IS BEING INVESTIGATED THROUGH CAPA.

Description of Event or Problem · 1

A CUSTOMER REPORTED TO BAXTER (B)(4) A UROMATIC EASY FLOW UNI-SET IN WHICH A LEAK WAS OBSERVED FROM AN UNKNOWN LOCATION. THE ALLEGED DEFECT WAS OBSERVED AFTER THE SET WAS ATTACHED TO THE SOLUTION BAG, BUT BEFORE PATIENT ATTACHMENT. THEREFORE, THERE WERE NO REPORTS OF PATIENT INVOLVEMENT, PATIENT INJURY, MEDICAL INTERVENTION, OR ADVERSE EVENTS ASSOCIATED WITH THIS REPORT. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ACCESS SYSTEM, IRRIGATION, UROLOGICAL LJH BAXTER HEALTHCARE - MALTA 11F08V757

Patients

Seq Age Sex Outcome Treatment
1