CD HORIZON® SOLERA® VOYAGER¿ SPINAL SYSTEM
Report
- Report Number
- 3003120897-2026-00094
- Event Type
- Malfunction
- Date Received
- January 20, 2026
- Date of Event
- December 29, 2025
- Report Date
- April 6, 2026
- Manufacturer
- MEDTRONIC SOFAMOR DANEK USA, INC.
- Product Code
- HWE
- UDI-DI
- 00643169754584
- PMA / PMN Number
- K170679
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.
H3: PRODUCT ANALYSIS OF PART # 6550017, LOT # K24F1188 VISUAL INSPECTION CONFIRMED THE END OF THE TAB EXTENDER THAT INTERFACES WITH THE BREAK OFF SCREW HAS BEEN BENT. THE DAMAGE TO THE EXTENDER IS CONSISTENT WITH OVERLOAD. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.
INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAVING SPINAL THERAPY FOR TH12 DISH. IT WAS REPORTED THAT DURING THE FINAL TAB BREAKING AFTER V5 WAS FULLY TIGHTENED, FORCE WAS APPLIED TO BREAK THE TAB BEFORE THE TAB BREAKER WAS FULLY INSERTED, RESULTING IN THE TAB BECOMING BENT. THERE WERE NO PATIENT SYMPTOMS REPORTED. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 176096 | CD HORIZON® SOLERA® VOYAGER¿ SPINAL SYSTEM | INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED M | HWE | MEDTRONIC SOFAMOR DANEK USA, INC. | 6550017 | K24F1188 | 00643169754584 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |