FDA Adverse Event Malfunction Summary report: N

VIDEO URETEROSCOPE, 8.5 FR. X 700MM

MDR report key: 24078822 · Received January 15, 2026

Report

Report Number
1221826-2026-00056
Event Type
Malfunction
Date Received
January 15, 2026
Date of Event
December 22, 2025
Report Date
February 25, 2026
Manufacturer
KARL STORZ SE & CO.KG
Product Code
FGB
PMA / PMN Number
K141250
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

CONCLUSION SUMMARY: NO IMAGE AND NO LIGHT. FLUID INVASION IN HANDLE HOUSING. LEAK FROM HANDLE HOUSING. NICKS ON DISTAL TIP. ROOT CAUSE OF HANDLE HOUSING LEAK WHICH CAUSED NO IMAGE AND NO LIGHT OUTPUT WAS IDENTIFIED ON ECM100007630 (EFFECTIVE 11/18/2024) AS A MANUFACTURING CHANGE TO THE HOUSING. KSE IS MONITORING THIS DEFECT WHICH IS CURRENTLY IN CONTROL. THIS EVENT IS FILED UNDER INTERNAL COMPLAINT ID (B)(4).

Additional Manufacturer Narrative · 0

THE DEVICE WILL BE FORWARDED TO THE MANUFACTURER FOR FURTHER INVESTIGATION. SHOULD RELEVANT ADDITIONAL INFORMATION / INVESTIGATION RESULTS BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED UNSOLICITED. THIS EVENT IS FILED UNDER INTERNAL COMPLAINT ID: (B)(4).

Description of Event or Problem · 0

IT WAS REPORTED THAT NO VIDEO IMAGE ON MONITOR. THERE WAS A DELAY OF LESS THAN 10 MIN TO RETRIEVE ANOTHER SCOPE FROM STERILE PROCESSING DEPARTMENT. NO NEGATIVE IMPACT IN STATE OF HEALTH REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
140654 VIDEO URETEROSCOPE, 8.5 FR. X 700MM VIDEO URETEROSCOPE, 8.5 FR. X 700MM FGB KARL STORZ SE & CO.KG R11278VSUEK

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown