FDA Adverse Event Injury Summary report: N

GMK REVISION TOTAL KNEE SYSTEM

MDR report key: 24078373 · Received January 15, 2026

Report

Report Number
3005180920-2026-00036
Event Type
Injury
Date Received
January 15, 2026
Date of Event
January 6, 2026
Report Date
January 15, 2026
Manufacturer
MEDACTA INTERNATIONAL SA
Product Code
JWH
UDI-DI
07630030819391
PMA / PMN Number
K123721
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

BATCH REVIEWS PERFORMED ON 07-JAN-26. GMK-REVISION 02.07.0684R REVISION TIBIAL TRAY SIZE 4 RIGHT (K123721) LOT. 189502: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 20-FEB-2019 EXPIRATION DATE: 2024-02-07. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN SOLD WITH NO SIMILAR REPORTED EVENT DURING THE PERIOD OF REVIEW. GMK-REVISION 02.07.2405R FEMUR REVISION PS CEMENTED S.5R (K102437) LOT. 1900614: (B)(4) ITEMS MANUFACTURED AND RELEASED ON 17-JUL-2019 EXPIRATION DATE: 2024-07-06. NO ANOMALIES FOUND RELATED TO THE PROBLEM. TO DATE, (B)(4) ITEMS OF THE SAME LOT HAVE BEEN SOLD WITH NO SIMILAR REPORTED EVENT DURING THE PERIOD OF REVIEW. ROOT CAUSE:BASED ON THE INFORMATION AVAILABLE NO DEFINITIVE ROOT CAUSE CAN BE ESTABLISHED, WHILE THERE IS NO INDICATION THAT ANY POTENTIAL ISSUE WITH THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT, AND THE DEVICE HISTORY RECORD REVIEW DOES NOT INDICATE ANY POTENTIAL MANUFACTURING RELATED ISSUE.

Description of Event or Problem · 0

THE PATIENT HAD PRIMARY RIGHT KNEE SURGERY USING COMPETITOR PRODUCTS. ON (B)(6) 2019, THE PATIENT WAS REVISED FROM THE COMPETITOR IMPLANTS TO MEDACTA IMPLANTS. PRESENTLY, ON (B)(6) 2026, THE PATIENT CAME IN DUE TO INSTABILITY AS A RESULT OF THE LOOSENING IMPLANTS OF THE TIBIA AND FEMUR. THE SURGEON REVISED THE FEMUR, TRAY, AND INSERT TO A GMK-HINGE. THE SURGERY WAS COMPLETED SUCCESSFULLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
145796 GMK REVISION TOTAL KNEE SYSTEM REVISION TIBIAL TRAY SIZE 4 RIGHT JWH MEDACTA INTERNATIONAL SA 02.07.0684R 189502 07630030819391

Patients

Seq Age Sex Outcome Treatment
1 67 YR Male Required Intervention