ERBE FLEXIBLE CRYOPROBE
Report
- Report Number
- 9610614-2026-00010
- Event Type
- Malfunction
- Date Received
- January 14, 2026
- Date of Event
- December 15, 2025
- Report Date
- January 14, 2026
- Manufacturer
- ERBE ELEKTROMEDIZIN GMBH
- Product Code
- GEH
- PMA / PMN Number
- K190651
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- OTHER
Narratives
THE CRYOPROBE WAS EVALUATED BY ERBE USA ON 2026-01-14. THE VISUAL EXAMINATION REVEALED A SLIT OR CUT IN THE WHITE TUBING APPROXIMATELY 34MM LONG AND LOCATED 32.5CM FROM THE DISTAL END OF THE WHITE TUBING. THE CLEAR TUBE WAS PULLED OUT FROM THE CONNECTOR THROUGH THE SLIT IN THE WHITE TUBE. THEN AT THE DIRECTION OF THE MANUFACTURER, ERBE ELEKTROMEDIZIN GMBH (ERBE GERMANY), THE CRYOPROBE'S CONNECTOR SECTION WAS DISASSEMBLED AS INSTRUCTED. GLUE BETWEEN THE CONNECTOR AND THE CLEAR PLASTIC TUBING ON THE HIGH-PRESSURE SIDE OF THE CRYOPROBE WAS FOUND TO BE INSUFFICIENT. THE HIGH-PRESSURE TUBING WAS ALSO PUSHED OUT OF THE CONNECTOR AND THE BLUE O-RING WAS STILL PRESENT. ASSESSMENT BY ERBE GERMANY. ERBE HAS BEEN MADE AWARE OF ISOLATED CASES WHERE THE OUTER WHITE TUBING OF CRYOPROBES RUPTURE. AFTER A THOROUGH INVESTIGATION, IT HAS BEEN CONCLUDED THAT A VERY LOW NUMBER OF ISOLATED INCIDENTS HAVE OCCURRED INVOLVING THE REPORTED FAILURE MODE. MORE SPECIFICALLY, LESS THAN (B)(4) OF CRYOPROBES MANUFACTURED DURING THE TIMEFRAME WHICH THE PROBLEM WAS DISCOVERED HAVE RUPTURED. IN THESE RARE CASES, THE FIXATION OF THE GAS INLET INSIDE THE PROBE CONNECTOR HAD LOOSENED, ALLOWING GAS TO FLOW INTO THE RETURN LINE. SINCE THE RETURN LINE IS NOT DESIGNED FOR THIS VOLUME OF GAS, THE TUBING COULD RUPTURE. SINCE DISCOVERING THE ISSUE, THE MANUFACTURING PROCESS HAS BEEN IMPROVED BY STABILIZING THE ADHESIVE APPLICATION AND ADDING ADDITIONAL VISUAL INSPECTION STEPS TO MONITOR THE GLUING OF EACH CRYOPROBE. ALL OF THESE MEASURES HAVE BEEN IMPLEMENTED TO MINIMIZE/ELIMINATE THE ISSUE. THE CUSTOMER IS BEING MADE AWARE OF THE FINDINGS. ERBE USA, INC. IS CLOSING THE FILE ON THIS EVENT.
IT WAS REPORTED THAT AN INCIDENT OCCURRED WITH A FLEXIBLE CRYOPROBE DURING A LUNG CRYOBIOPSY. THE CRYOPROBE WAS USED WITH AN ERBE CRYOSURGICAL UNIT (MODEL ERBECRYO 2, PART NUMBER: 10402-000, SERIAL NUMBER: NOT PROVIDED). NO OTHER INFORMATION WAS PROVIDED REGARDING ANY OTHER ACCESSORY EMPLOYED DURING THE INCIDENT. THE CRYOSURGICAL UNIT'S SETTING WAS 54 BAR. PER THE COMPLAINANT, ON THE SECOND BIOPSY, THE CRYOPROBE BURST AND PART OF IT WAS SEEN DAMAGED WHERE IT CONNECTS TO THE UNIT. THERE WAS NO REPORT OF ANY USER OR PATIENT INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 128250 | ERBE FLEXIBLE CRYOPROBE | CRYOPROBE | GEH | ERBE ELEKTROMEDIZIN GMBH |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 50 YR | Unknown |