FDA Adverse Event Malfunction Summary report: N

COBAS C 503 ANALYTICAL UNIT

MDR report key: 24066727 · Received January 14, 2026

Report

Report Number
1823260-2026-00174
Event Type
Malfunction
Date Received
January 14, 2026
Date of Event
December 25, 2025
Report Date
March 20, 2026
Manufacturer
ROCHE DIAGNOSTICS
Product Code
JJE
PMA / PMN Number
K191899
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

CALIBRATION AND QC WERE PASSING AT THE TIME OF THE EVENT. THERE WAS NO INDICATION OF A PERFORMANCE ISSUE WITH THE REAGENT OR ANALYZER. THE CUSTOMER'S CENTRIFUGE TIME MAY BE TOO SHORT, AND THE SPEED MAY BE TOO HIGH PER THE MANUFACTURER'S RECOMMENDED GUIDELINES. THIS CAN BE CONSISTENT WITH PRE-ANALYTICAL ISSUES. IN THE ALARM TRACE REPORT, THERE WERE ALARMS FOR ABNORMAL ASPIRATION, WHICH CAN ALSO BE CONSISTENT WITH A PRE-ANALYTICAL ISSUE. A FIELD SERVICE ENGINEER (FSE) EVALUATED THE INSTRUMENT AND DID NOT FIND A ROOT CAUSE FOR THE CUSTOMER'S ISSUE. THE INSTRUMENT PERFORMANCE WAS VERIFIED, AND THE ISSUE APPEARS TO BE RESOLVED. THE INVESTIGATION WAS UNABLE TO DETERMINE A DIRECT ROOT CAUSE. THE CUSTOMER HAS NOT REPORTED ANY FURTHER ISSUES SINCE THE SERVICE VISIT WAS COMPLETE.

Additional Manufacturer Narrative · 0

THE GLUCOSE HK GEN.3 REAGENT LOT NUMBER IS 882312; THE EXPIRATION DATE WAS NOT PROVIDED. THE INVESTIGATION IS ONGOING.

Description of Event or Problem · 0

THERE WAS AN ALLEGATION OF A QUESTIONABLE GLUCOSE HK GEN.3 RESULT FROM THE COBAS C 503 ANALYTICAL UNIT FOR ONE PATIENT. THE INITIAL RESULT WAS 15 MG/DL, AND THE REPEAT RESULT ON ANOTHER C 503 WAS 143 MG/DL. THE REPEAT RESULT WAS DEEMED TO BE CORRECT. THE CUSTOMER FOUND THE INITIAL RESULT TO BE QUESTIONABLE AND REPEATED IT. THE INITIAL RESULT WAS NOT RELEASED OUTSIDE OF THE LABORATORY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
129254 COBAS C 503 ANALYTICAL UNIT CLINICAL CHEMISTRY ANALYZER JJE ROCHE DIAGNOSTICS

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown