FDA Adverse Event Malfunction Summary report: N

QUICK SET

MDR report key: 24065847 · Received January 14, 2026

Report

Report Number
3003442380-2025-12679
Event Type
Malfunction
Date Received
January 14, 2026
Date of Event
December 25, 2025
Report Date
March 18, 2026
Manufacturer
UNOMEDICAL DEVICES S.A. DE C.V.
Product Code
FPA
UDI-DI
05705244017573
PMA / PMN Number
K160648
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

ADDITIONAL INFORMATION - THIS MDR IS BEING SUBMITTED TO INCLUDE THE BELOW: H6: INVESTIGATION RESULTS UNDER TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, INVESTIGATION CONCLUSIONS. H11: INVESTIGATION SUMMARY. COMPLAINT INVESTIGATION RESULTS: ELECTRONIC QUALITY MANAGEMENT SYSTEM (EQMS) SEARCH: A QUERY WAS RUN IN THE EQMS ON 18-MAR-2026 AGAINST "LOT NUMBER 6011624 AND SIMILAR MALFUNCTION CODE(S): TUBING DAMAGE - SIGNIFICANT / EXTENSIVE TUBING IS SPLIT (ALONG THE LENGTH OF THE TUBING) OR HAS PINHOLES THE REVIEW CONFIRMED THAT LOT 6011624 AND THE IDENTIFIED FAILURE MODE ARE NOT ASSOCIATED WITH ANY NONCONFORMING REPORTS (NCRS) OR CORRECTIVE AND PREVENTIVE ACTION (CAPAS) OF THE SAME OR SIMILAR NATURE. SIMILAR COMPLAINTS SEARCH: A QUERY WAS RUN IN THE EQMS ON 18-MAR-2026 AGAINST "LOT NUMBER" CRITERIA EQUAL 6011624 AND SIMILAR MALFUNCTION CODES TUBING DAMAGE - SIGNIFICANT / EXTENSIVE TUBING IS SPLIT (ALONG THE LENGTH OF THE TUBING) OR HAS PINHOLES THE COUNT OF COMPLAINT IS 1. THE COMPLAINT NUMBER IS (B)(4). DEVICE HISTORY RECORD (DHR) REVIEW: THE LOT 6011624 WAS MANUFACTURED ACCORDING TO THE WORK INSTRUCTION (WI) VERSION 82 AND MANUFACTURED IN THE LINE 12 ON 12/FEB/2025 WITH A TOTAL OF (B)(4) UNITS. THE SUB-ASSEMBLY, GLUING OF TUBING OF THE LOT 5B00438 WAS MANUFACTURED ACCORDING TO THE WI VERSION 42 AND MANUFACTURED IN THE MACHINE 04, 08, ON 08/FEB/2025, WITH A TOTAL OF (B)(4) UNITS. THE SUB-ASSEMBLY, GLUING OF TUBING OF THE LOT 5B01648 WAS MANUFACTURED ACCORDING TO THE WI VERSION 42 AND MANUFACTURED IN THE MACHINE 04, 08, ON 02/DEC/2024, WITH A TOTAL OF (B)(4) UNITS. THE SUB-ASSEMBLY, GLUING OF TUBING OF THE LOT 5B000439 WAS MANUFACTURED ACCORDING TO THE WI VERSION 42 AND MANUFACTURED IN THE MACHINE 04, 08, ON 11/FEB/2025, WITH A TOTAL OF (B)(4) UNITS. IN THE TEST 9 IS FOUND 1SET NEEDLE BENT, EXTENDED SAMPLING PLAN ACCEPTABLE. THEREFORE, THE OVERALL REVIEW OF THE DHR CONFIRMS THAT ALL REQUIRED PROCESS-RELATED TESTS WERE COMPLETED AND MET THE APPLICABLE REQUIREMENTS. NO DEVIATIONS WERE IDENTIFIED, AND NO MAINTENANCE EVENTS WERE RECORDED IN RELATION TO THE COMPLAINT CODE. CONCLUSION: DHR REVIEW IDENTIFIED MINOR FINDINGS. THEY WERE MANAGED ACCORDING TO PROCEDURE AND DID NOT IMPACT COMPLIANCE; NO ISSUES NOTED. VISUAL EVIDENCE REVIEW: NO PHOTO WAS PROVIDED TO SUPPORT VISUAL CONFIRMATION OF THE REPORTED ISSUE. CONCLUSION: UNABLE TO PERFORM VISUAL VERIFICATION; ASSESSMENT BASED ON AVAILABLE DOCUMENTATION ONLY. CAPA DETERMINATION ASSESSMENT - CONCLUSION SUMMARY OF COMPLAINT INVESTIGATION: BASED ON THE INVESTIGATION, NO FURTHER INVESTIGATION IS REQUIRED. THE RECORD WILL BE CLOSED AND MONITORED THROUGH TRACKING AND TRENDING PER WI (MONTHLY TRIPS AND ALERTS). CONCLUSION SUMMARY OF COMPLAINT INVESTIGATION: AS A RESULT OF THE FOLLOWING; A COMPREHENSIVE REVIEW WAS CONDUCTED, INCLUDING EQMS QUERIES, SIMILAR COMPLAINT SEARCHES, DEVICE HISTORY RECORD REVIEW, VISUAL EVIDENCE ASSESSMENT, AND CAPA DETERMINATION. NO NCRS OR CAPAS OF THE SAME OR SIMILAR NATURE WERE FOUND FOR LOT 6011624 AND RELATED MALFUNCTION CODES FOR LEAK ISSUES. ONE COMPLAINT WAS IDENTIFIED FOR THIS LOT, BUT NO TREND OR SYSTEMIC ISSUE WAS DETECTED. THE MANUFACTURING RECORDS CONFIRMED THAT THE LOT WAS PRODUCED IN COMPLIANCE WITH ALL REQUIREMENTS, WITH NO DEVIATIONS OR MAINTENANCE EVENTS NOTED. NO PHOTO EVIDENCE WAS PROVIDED, SO THE ASSESSMENT WAS BASED ON DOCUMENTATION ONLY. BASED ON THESE RESULTS, NO MANUFACTURING OR QUALITY ISSUES WERE IDENTIFIED, AND NO FURTHER INVESTIGATION IS REQUIRED. THE RECORD WILL BE CLOSED AND MONITORED THROUGH ROUTINE TRACKING AND TRENDING.

Description of Event or Problem · 0

TO DATE NO ADDITIONAL PATIENT OR EVENT DETAILS HAVE BEEN RECEIVED.

Description of Event or Problem · 0

REFERENCE NUMBER (B)(4). EVENT OCCURRED IN THE UNITED STATES. IT WAS REPORTED THAT THE PATIENT FACED INFUSION SET TUBING KINKED EVENT ON (B)(6) 2025 WHICH LED TO HYPERGLYCAEMIA WITH BLOOD GLUCOSE VALUE OF 115 MG/DL. THE PATIENT GOT TREATED WITH MANUAL INJECTIONS. NO FURTHER INFORMATION AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
34321 QUICK SET UNO QUICK-SET 60/9 SC1 MECA FPA UNOMEDICAL DEVICES S.A. DE C.V. MMT-397A 6011624 05705244017573

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown