FDA Adverse Event Malfunction Summary report: N

I-STAT TROPONIN (HS-TNI ) CARTRIDGE

MDR report key: 24059282 · Received January 13, 2026

Report

Report Number
2245578-2026-00010
Event Type
Malfunction
Date Received
January 13, 2026
Date of Event
December 24, 2025
Report Date
March 2, 2026
Manufacturer
ABBOTT POINT OF CARE INC.
Product Code
MMI
PMA / PMN Number
K240984
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NE, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

APOC INCIDENT: (B)(4). THE INVESTIGATION WAS COMPLETED ON 24-FEB-2026. A REVIEW OF THE DEVICE HISTORY RECORD (DHR) CONFIRMED THE CARTRIDGE LOT MET FINISHED GOODS (FG) RELEASE CRITERIA. A DEFICIENCY HAS BEEN IDENTIFIED FOR HS-TNI CARTRIDGE LOT A25227. WHILE RETAINED CARTRIDGE TESTING MET THE ACCEPTANCE CRITERIA FOR THE DETECTED ERROR (DT) AND UNDETECTED ERROR (UDT) RATES OUTLINED IN APPENDIX 1 OF Q04.01.003 REV. AP (PRODUCT COMPLAINT LEVEL 2 AND LEVEL 3 INVESTIGATION PROCEDURE), THE FG RELEASE SPECIFICATION FOR STANDARD DEVIATION (SD) PER TP-0651 REV. BI (CARTRIDGE FINISHED GOODS TEST SPECIFICATION) WAS NOT MET. ROOT CAUSE WILL BE INVESTIGATED IN QUALITY RECORD (QR) 1100932.

Additional Manufacturer Narrative · 0

APOC INCIDENT # (B)(4), APOC LABELING WILL BE EVALUATED DURING THE INVESTIGATION AS PERTAINING TO THE EVENT.

Description of Event or Problem · 0

ON (B)(6) 2025, ABBOTT POINT OF CARE WAS CONTACTED BY A CUSTOMER REGARDING I-STAT TROPONIN CARTRIDGES THAT YIELDED DISCREPANT FALSE POSITIVE RESULTS ON A 83-YEAR-OLD MALE PATIENT PRESENTED WITH NEAR SYNCOPE EPISODE AND NAUSEA. THERE WAS NO ADDITIONAL PATIENT INFORMATION. RETURN PRODUCT IS NOT AVAILABLE FOR INVESTIGATION. (B)(6). CUTPFFS FOR BOTH METHODS: WOMEN > 36 PG/ML MEN > 60 PG/ML = PROBABLE. THERE ARE NO INJURIES ASSOCIATED WITH THIS EVENT. AT THIS TIME THERE IS NO REASON TO SUSPECT A MALFUNCTION EXISTS. THE INVESTIGATION IS UNDERWAY. (B)(6)

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
478409 I-STAT TROPONIN (HS-TNI ) CARTRIDGE HS-TNI CARTRIDGE MMI ABBOTT POINT OF CARE INC. NA A25227

Patients

Seq Age Sex Outcome Treatment
1 84 YR Male