FDA Adverse Event Malfunction Summary report: N

RATCHETING SCREWDRIVER HANDLE

MDR report key: 24057347 · Received January 13, 2026

Report

Report Number
8030965-2026-00448
Event Type
Malfunction
Date Received
January 13, 2026
Date of Event
December 24, 2025
Manufacturer
SYNTHES GMBH
Product Code
HXX
UDI-DI
07611819895164
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF THE INFORMATION IS UNKNOWN, NOT AVAILABLE OR DOES NOT APPLY, THE SECTION/FIELD OF THE FORM IS LEFT BLANK. H11 ADDITIONAL NARRATIVE: H6 COMPONENT CODES: MOST RELEVANT COMPONENT CODE IS G07002 (APPROPRIATE TERM/CODE NOT AVAILABLE) TO CAPTURE NO FINDINGS AVAILABLE DUE TO NO PRODUCT RETURNED. H6: THE PRODUCT WAS NOT RETURNED TO DEPUY SYNTHES; HOWEVER, PHOTOS WERE PROVIDED FOR REVIEW. THE DEVICE ASSOCIATED WITH THIS REPORT WAS NOT RETURNED TO DEPUY SYNTHES FOR EVALUATION. THE PHOTOGRAPHS ATTACHED WERE REVIEWED, HOWEVER THE PROVIDED EVIDENCE WAS NOT SUFFICIENT TO CONFIRM THE REPORTED EVENT OF "UNABLE TO ASSEMBLE/DISASSEMBLE (2+ DEVICES)" . DEVICE INTERACTION ISSUES CANNOT BE EVALUATED THROUGH A PHOTO INVESTIGATION. SINCE THE DEVICE WAS NOT RETURNED, A DIMENSIONAL INSPECTION CANNOT BE PERFORMED. THE OVERALL COMPLAINT WAS UNCONFIRMED AS THE PHOTOGRAPHS ATTACHED DON¿T HAVE ANY EVIDENCE TO DRAW A CONCLUSION ABOUT THE REPORTED EVENT. BASED ON THE INVESTIGATION FINDINGS, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. THERE IS NO INDICATION THAT A DESIGN OR MANUFACTURING ISSUE HAS CAUSED THE COMPLAINT CONDITION. AS PART OF DEPUY SYNTHES QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. ADDITIONAL MONITORING FOR ANY POTENTIAL SAFETY SIGNALS WILL BE CONDUCTED THROUGH COMPLAINT TRENDING AND OTHER POST-MARKET SAFETY SURVEILLANCE ACTIVITIES. DEVICE HISTORY LOT: PART NUMBER: 311.023, LOT NUMBER: T162466, MANUFACTURING SITE: TUTTLINGEN, RELEASE TO WAREHOUSE DATE: 24-JUL-2018. A REVIEW OF THE DEVICE HISTORY RECORDS WAS PERFORMED FOR THE FINISHED DEVICE LOT NUMBER, AND NO NON-CONFORMANCES WERE IDENTIFIED. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 0

THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803 (AND/OR PART 4, AS APPLICABLE). THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY DEPUY SYNTHES, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL REPORT, A FOLLOW-UP REPORT WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 0

DURING THE SURGERY, WHILE THE DOCTOR WAS USING THE SCREWDRIVER, A DIFFICULTY AROSE IN DISENGAGING THE HANDLE FROM THE BLADE. AN ATTEMPT WAS MADE TO REATTACH THE HANDLE TO THE BLADE, BUT A PROPER CONNECTION COULD NOT BE ACHIEVED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
56328 RATCHETING SCREWDRIVER HANDLE SCREWDRIVER HXX SYNTHES GMBH T162466 07611819895164

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown