FDA Adverse Event Malfunction Summary report: N

GENTLEPOWER LUX CONTRA ANGLE 25LPA

MDR report key: 2405703 · Received December 7, 2011

Report

Report Number
3003637274-2011-00067
Event Type
Malfunction
Date Received
December 7, 2011
Date of Event
November 9, 2011
Report Date
November 10, 2011
Manufacturer
KAVO DENTAL GMBH
Product Code
EFB
PMA / PMN Number
K073478
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UT, US
Reporter Occupation
DENTIST

Narratives

Additional Manufacturer Narrative · 1

A) DURING A STANDARD TREATMENT, AN ELECTRICAL DENTAL HANDPIECE GOT HOT AND BURNED THE PT ON THE TONGUE AND LOWER LIP. TONGUE HAD A VERY SMALL BURN, LIP WAS BURNED TO THE SIZE OF THE BACK CAP. PT WAS TOLD TO TAKE OTC-IBUPROFEN AND WAS INSTRUCTED TO APPLY ICE TO LIP AS NEEDED. PT WAS CONTACTED ONE DAY AFTER AND SHE STATED THAT SHE WAS FEELING WELL. B) THE ANALYSIS OF THE HANDPIECE DID SHOW THAT IT HAD A MEDIUM DEBRIS LEVEL. THE BEARINGS HAVE BEEN GRINDING, WOBBLING AND FALLING APART. THE HEAD HEATED UP DURING TEST RUN OF THE EVAL. THE USER INSTRUCTION REQUIRES A TEST RUN BEFORE EACH USE AND INFORMS THAT A HANDPIECE MUSTN'T BE USED IF IT RUNS OUT OF SPECIFICATIONS. C) REPORTED DUE TO THE 2 YEAR PRESUMPTION RULE.

Description of Event or Problem · 1

DURING A STANDARD TREATMENT, AN ELECTRICAL DENTAL HANDPIECE GOT HOT AND BURNED THE PT ON TONGUE AND LOWER LIP. TONGUE HAD A VERY SMALL BURN, LIP WAS BURNED TO THE SIZE OF THE BACK CAP. PT WAS TOLD TO TAKE OTC-IBUPROFEN AND WAS INSTRUCTED TO APPLY ICE TO LIP AS NEEDED. PT WAS CONTACTED ONE DAY AFTER AND SHE STATED THAT SHE WAS FEELING WELL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 GENTLEPOWER LUX CONTRA ANGLE 25LPA DENTAL HANDPIECE EFB KAVO DENTAL GMBH 25LPA

Patients

Seq Age Sex Outcome Treatment
1 28 YR