BD VACUTAINER® ECLIPSE¿ BLOOD COLLECTION NEEDLE PRE-ATTACHED HOLDER
Report
- Report Number
- 1024879-2025-02420
- Event Type
- Malfunction
- Date Received
- January 13, 2026
- Date of Event
- December 16, 2025
- Report Date
- February 10, 2026
- Manufacturer
- BECTON DICKINSON & CO (FRANKLIN LAKES)
- Product Code
- JKA
- UDI-DI
- 50382903686501
- PMA / PMN Number
- K243207
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FI
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
INVESTIGATION SUMMARY: BD DID NOT RECEIVE ANY SAMPLES OR PHOTOS FOR INVESTIGATION. BASED ON A REVIEW OF THE DEVICE HISTORY RECORD FOR THE INCIDENT LOTS, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET. THIS COMPLAINT HAS NOT BEEN CONFIRMED FOR THE LOTS 4330752 AND 5121520, FOR THE INDICATED FAILURE MODE: DEFECTIVE LOCKING MECHANISM. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. INFORMATION WILL BE CAPTURED ON TREND REPORTS AND MONITORED. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR THE IDENTIFICATION OF EMERGING TRENDS.
THERE WERE MULTIPLE 510K NUMBERS REPORTED TO BE INVOLVED. THE INFORMATION IS AS FOLLOWS: G.4. PMA / 510(K)#: K982541. A DEVICE EVALUATION IS ANTICIPATED BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
IT WAS REPORTED THAT WHEN USING THE BD VACUTAINER® ECLIPSE¿ BLOOD COLLECTION NEEDLE PRE-ATTACHED HOLDER, TWENTY (20) UNITS EXHIBITED THE SAFETY MECHANISM SEPARATE FROM THE UNIT DURING ACTIVATION. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.
IT WAS REPORTED THAT WHEN USING THE BD VACUTAINER® ECLIPSE¿ BLOOD COLLECTION NEEDLE PRE-ATTACHED HOLDER, TWENTY (20) UNITS EXHIBITED THE SAFETY MECHANISM SEPARATE FROM THE UNIT DURING ACTIVATION. NO ADVERSE IMPACT WAS REPORTED REGARDING THE PATIENT OR USER.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 56310 | BD VACUTAINER® ECLIPSE¿ BLOOD COLLECTION NEEDLE PRE-ATTACHED HOLDER | TUBES, VIALS, SYSTEMS, SERUM SEPARATORS, BLOOD COLLECTION | JKA | BECTON DICKINSON & CO (FRANKLIN LAKES) | 5121520 | 50382903686501 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |