FDA Adverse Event Injury Summary report: N

INSET

MDR report key: 24051595 · Received January 13, 2026

Report

Report Number
3003442380-2025-18789
Event Type
Injury
Date Received
January 13, 2026
Date of Event
November 26, 2025
Report Date
January 13, 2026
Manufacturer
UNOMEDICAL DEVICES S.A. DE C.V.
Product Code
FPA
UDI-DI
05705244016620
PMA / PMN Number
K032854
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

ADDITIONAL INFORMATION - THIS MDR IS BEING SUBMITTED TO INCLUDE THE BELOW: H6: INVESTIGATION RESULTS UNDER TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, INVESTIGATION CONCLUSIONS. H11: INVESTIGATION SUMMARY COMPLAINT INVESTIGATION RESULTS: ELECTRONIC QUALITY MANAGEMENT SYSTEM (EQMS)SEARCH: A QUERY WAS RUN IN THE EQMS ON 12-JAN-2026 AGAINST "LOT NUMBER 6013375 AND SIMILAR MALFUNCTION CODE(S):SOFT CANNULA BENT/KINKED/CRIMPED AFTER REMOVAL - BLOCKAGE, SOFT CANNULA FOUND BENT UPON REMOVAL FROM INFUSION SITE, SOFT CANNULA FOUND KINKED UPON REMOVAL FROM INFUSION SITE, SOFT CANNULA FOUND CRIMPED UPON REMOVAL FROM INFUSION SITE, SOFT CANNULA AND INTRODUCER NEEDLE FOUND BENT/KINKED DURING INSERTION, UNABLE TO USE,SOFT CANNULA BENT (NO HARM). SOFT CANNULA BENT/KINKED/CRIMPED AFTER REMOVAL - REDUCED FLOW. THE REVIEW CONFIRMED THAT LOT 6013474 AND THE IDENTIFIED FAILURE MODE ARE NOT ASSOCIATED WITH ANY NCRS OR CORRECTIVE AND PREVENTIVE ACTION (CAPA) OF THE SAME OR SIMILAR NATURE. SIMILAR COMPLAINTS SEARCH: A QUERY WAS RUN IN THE EQMS ON 12-JAN-2026 AGAINST "LOT NUMBER" CRITERIA EQUAL 6013375 AND SIMILAR MALFUNCTION CODES SOFT CANNULA BENT/KINKED/CRIMPED AFTER REMOVAL - BLOCKAGE, SOFT CANNULA FOUND BENT UPON REMOVAL FROM INFUSION SITE, SOFT CANNULA FOUND KINKED UPON REMOVAL FROM INFUSION SITE, SOFT CANNULA FOUND CRIMPED UPON REMOVAL FROM INFUSION SITE, SOFT CANNULA AND INTRODUCER NEEDLE FOUND BENT/KINKED DURING INSERTION, UNABLE TO USE,SOFT CANNULA BENT (NO HARM). SOFT CANNULA BENT/KINKED/CRIMPED AFTER REMOVAL - REDUCED FLOW. THE COUNT OF COMPLAINT IS 1. THE COMPLAINT NUMBERS ARE (B)(4). DHR REVIEW: THE LOT 6013375 WAS MANUFACTURED ACCORDING TO THE WORK INSTRUCTION (WI) VERSION 123 AND MANUFACTURED IN THE LINE 08, ON 21-MAY-2025, WITH A TOTAL OF (B)(4) UNITS. THE SUB-ASSEMBLY. GLUING OF TUBING OF THE LOT 5E00711 WAS MANUFACTURED ACCORDING TO THE FORM-001865 VERSION 03 AND MANUFACTURED IN THE GLUING MACHINE ITL03, ON 13-MAY-2025, WITH A TOTAL OF (B)(4) UNITS. THE SUB-ASSEMBLY. GLUING OF TUBING OF THE LOT 5D01203 WAS MANUFACTURED ACCORDING TO THE WI VERSION 42 AND MANUFACTURED IN THE GLUING MACHINE ITL03, ON 23-APR-2025, WITH A TOTAL OF (B)(4) UNITS. THE SUB-ASSEMBLY. GLUING OF TUBING OF THE LOT 5E00712 WAS MANUFACTURED ACCORDING TO THE WI VERSION 42 AND MANUFACTURED IN THE GLUING MACHINE ITL03, ON 17-MAY-2025, WITH A TOTAL OF (B)(4) UNITS. THE DEVICE HISTORY RECORD (DHR) WAS REVIEWED IN ACCORDANCE WITH APPLICABLE PROCEDURES. ALL REQUIRED IN-PROCESS AND FINAL TESTS WERE COMPLETED AND MET SPECIFIED REQUIREMENTS. NO DEVIATIONS WERE IDENTIFIED, AND NO MAINTENANCE EVENTS WERE RECORDED THAT RELATE TO THE COMPLAINT CODE. CONCLUSION: DHR REVIEW SUPPORTS COMPLIANCE WITH MANUFACTURING AND QUALITY REQUIREMENTS; NO ISSUES NOTED. VISUAL EVIDENCE REVIEW: NO PHOTO WAS PROVIDED TO SUPPORT VISUAL CONFIRMATION OF THE REPORTED ISSUE. CONCLUSION: UNABLE TO PERFORM VISUAL VERIFICATION; ASSESSMENT BASED ON AVAILABLE DOCUMENTATION ONLY. RETAIN SAMPLES TESTING: RETAIN SAMPLES FROM THE RELEVANT LOT WERE REQUESTED AND TESTED IN ACCORDANCE WITH APPROVED PROCEDURES: WI - GUIDANCE FOR VISUAL TESTING FOR COMPLAINTS AREA VERSION 3: ALL 3 SAMPLES TESTED PASSED VISUAL INSPECTION. WI- GUIDANCE FOR FUNCTIONAL TESTING 1 AIR FLOW TEST FOR COMPLAINTS AREA VERSION 2: ALL 3 SAMPLES TESTED PASSED FUNCTIONAL TESTING FOR THE REPORTED MALFUNCTION CODE SOFT CANNULA BENT/KINKED/CRIMPED AFTER REMOVAL - BLOCKAGE. CONCLUSION: TESTING DID NOT CONFIRM THE REPORTED ISSUE; NO NONCONFORMANCE IDENTIFIED. CONCLUSION SUMMARY OF COMPLAINT INVESTIGATION: BASED ON THE INVESTIGATION, NO FURTHER INVESTIGATION IS REQUIRED. THE RECORD WILL BE CLOSED AND MONITORED THROUGH TRACKING AND TRENDING PER WI-001031 (MONTHLY TRIPS AND ALERTS). AS A RESULT OF THE FOLLOWING A COMPREHENSIVE REVIEW WAS CONDUCTED, INCLUDING EQMS QUERIES, SIMILAR COMPLAINT SEARCHES, DEVICE HISTORY RECORD REVIEW, VISUAL EVIDENCE ASSESSMENT, AND CAPA DETERMINATION. NO NCRS OR CAPAS OF THE SAME OR SIMILAR NATURE WERE FOUND FOR LOT# 6013375 AND RELATED MALFUNCTION CODES. ONE COMPLAINT WAS IDENTIFIED FOR THIS LOT; HOWEVER, NO TREND OR SYSTEMIC ISSUE WAS DETECTED. THE MANUFACTURING RECORDS CONFIRMED THAT THE LOT WAS PRODUCED IN COMPLIANCE WITH ALL REQUIREMENTS, WITH NO DEVIATIONS OR MAINTENANCE EVENTS NOTED. SAMPLES WERE REQUESTED; HOWEVER, THE CUSTOMER CONFIRMED THAT NO SAMPLES WERE AVAILABLE FOR ANALYSIS. CONSEQUENTLY, AN INVESTIGATION WAS CONDUCTED USING REFERENCE SAMPLES, AND NO FAILURES RELATED TO THE COMPLAINT WERE IDENTIFIED. NO PHOTO EVIDENCE WAS PROVIDED, SO THE ASSESSMENT WAS BASED ON DOCUMENTATION AND REFERENCE SAMPLE ANALYSIS ONLY. BASED ON THESE RESULTS, NO MANUFACTURING OR QUALITY ISSUES WERE IDENTIFIED. THE RECORD WILL BE CLOSED AND MONITORED THROUGH ROUTINE TRACKING AND TRENDING.

Additional Manufacturer Narrative · 0

ADDITIONAL INFORMATION - THIS MDR IS BEING SUBMITTED TO INCLUDE THE BELOW: H6: INVESTIGATION RESULTS UNDER TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, INVESTIGATION CONCLUSIONS. H11: INVESTIGATION SUMMARY. COMPLAINT INVESTIGATION RESULTS: ELECTRONIC QUALITY MANAGEMENT SYSTEM (EQMS)SEARCH: A QUERY WAS RUN IN THE EQMS ON 12-JAN-2026 AGAINST "LOT NUMBER 6013375 AND SIMILAR MALFUNCTION CODE(S):SOFT CANNULA BENT/KINKED/CRIMPED AFTER REMOVAL - BLOCKAGE, SOFT CANNULA FOUND BENT UPON REMOVAL FROM INFUSION SITE, SOFT CANNULA FOUND KINKED UPON REMOVAL FROM INFUSION SITE, SOFT CANNULA FOUND CRIMPED UPON REMOVAL FROM INFUSION SITE, SOFT CANNULA AND INTRODUCER NEEDLE FOUND BENT/KINKED DURING INSERTION, UNABLE TO USE,SOFT CANNULA BENT (NO HARM). SOFT CANNULA BENT/KINKED/CRIMPED AFTER REMOVAL - REDUCED FLOW. THE REVIEW CONFIRMED THAT LOT 6013474 AND THE IDENTIFIED FAILURE MODE ARE NOT ASSOCIATED WITH ANY NCRS OR CORRECTIVE AND PREVENTIVE ACTION (CAPA) OF THE SAME OR SIMILAR NATURE. SIMILAR COMPLAINTS SEARCH: A QUERY WAS RUN IN THE EQMS ON 12-JAN-2026 AGAINST "LOT NUMBER" CRITERIA EQUAL 6013375 AND SIMILAR MALFUNCTION CODES SOFT CANNULA BENT/KINKED/CRIMPED AFTER REMOVAL - BLOCKAGE, SOFT CANNULA FOUND BENT UPON REMOVAL FROM INFUSION SITE, SOFT CANNULA FOUND KINKED UPON REMOVAL FROM INFUSION SITE, SOFT CANNULA FOUND CRIMPED UPON REMOVAL FROM INFUSION SITE, SOFT CANNULA AND INTRODUCER NEEDLE FOUND BENT/KINKED DURING INSERTION, UNABLE TO USE,SOFT CANNULA BENT (NO HARM). SOFT CANNULA BENT/KINKED/CRIMPED AFTER REMOVAL - REDUCED FLOW.THE COUNT OF COMPLAINT IS 1. THE COMPLAINT NUMBERS ARE (B)(4). DEVICE HISTORY RECORD (DHR) REVIEW: THE LOT 6013375 WAS MANUFACTURED ACCORDING TO THE WORK INSTRUCTION (WI) VERSION 123 AND MANUFACTURED IN THE LINE 08, ON 21- MAY-2025, WITH A TOTAL OF (B)(4) UNITS. THE SUB-ASSEMBLY. GLUING OF TUBING OF THE LOT 5E00711 WAS MANUFACTURED ACCORDING TO THE FORM-001865 VERSION 03 AND MANUFACTURED IN THE GLUING MACHINE ITL03, ON 13-MAY-2025, WITH A TOTAL OF (B)(4) UNITS. THE SUB-ASSEMBLY. GLUING OF TUBING OF THE LOT 5D01203 WAS MANUFACTURED ACCORDING TO THE WI VERSION 42 AND MANUFACTURED IN THE GLUING MACHINE ITL03, ON 23-APR-2025, WITH A TOTAL OF (B)(4) UNITS. THE SUB-ASSEMBLY. GLUING OF TUBING OF THE LOT 5E00712 WAS MANUFACTURED ACCORDING TO THE WI VERSION 42 AND MANUFACTURED IN THE GLUING MACHINE ITL03, ON 17-MAY-2025, WITH A TOTAL OF (B)(4) UNITS. THE DEVICE HISTORY RECORD (DHR) WAS REVIEWED IN ACCORDANCE WITH APPLICABLE PROCEDURES. ALL REQUIRED IN-PROCESS AND FINAL TESTS WERE COMPLETED AND MET SPECIFIED REQUIREMENTS. NO DEVIATIONS WERE IDENTIFIED, AND NO MAINTENANCE EVENTS WERE RECORDED THAT RELATE TO THE COMPLAINT CODE. CONCLUSION: DHR REVIEW SUPPORTS COMPLIANCE WITH MANUFACTURING AND QUALITY REQUIREMENTS; NO ISSUES NOTED. VISUAL EVIDENCE REVIEW: NO PHOTO WAS PROVIDED TO SUPPORT VISUAL CONFIRMATION OF THE REPORTED ISSUE. CONCLUSION: UNABLE TO PERFORM VISUAL VERIFICATION; ASSESSMENT BASED ON AVAILABLE DOCUMENTATION ONLY. RETAIN SAMPLES TESTING: RETAIN SAMPLES FROM THE RELEVANT LOT WERE REQUESTED AND TESTED IN ACCORDANCE WITH APPROVED PROCEDURES: WI - GUIDANCE FOR VISUAL TESTING FOR COMPLAINTS AREA VERSION 3: ALL 3 SAMPLES TESTED PASSED VISUAL INSPECTION. WI- GUIDANCE FOR FUNCTIONAL TESTING 1 AIR FLOW TEST FOR COMPLAINTS AREA VERSION 2: ALL 3 SAMPLES TESTED PASSED FUNCTIONAL TESTING FOR THE REPORTED MALFUNCTION CODE SOFT CANNULA BENT/KINKED/CRIMPED AFTER REMOVAL - BLOCKAGE. CONCLUSION: TESTING DID NOT CONFIRM THE REPORTED ISSUE; NO NONCONFORMANCE IDENTIFIED. CONCLUSION SUMMARY OF COMPLAINT INVESTIGATION: BASED ON THE INVESTIGATION, NO FURTHER INVESTIGATION IS REQUIRED. THE RECORD WILL BE CLOSED AND MONITORED THROUGH TRACKING AND TRENDING PER WI-001031 (MONTHLY TRIPS AND ALERTS). AS A RESULT OF THE FOLLOWING A COMPREHENSIVE REVIEW WAS CONDUCTED, INCLUDING EQMS QUERIES, SIMILAR COMPLAINT SEARCHES, DEVICE HISTORY RECORD REVIEW, VISUAL EVIDENCE ASSESSMENT, AND CAPA DETERMINATION. NO NCRS OR CAPAS OF THE SAME OR SIMILAR NATURE WERE FOUND FOR LOT 6013375 AND RELATED MALFUNCTION CODES. ONE COMPLAINT WAS IDENTIFIED FOR THIS LOT; HOWEVER, NO TREND OR SYSTEMIC ISSUE WAS DETECTED. THE MANUFACTURING RECORDS CONFIRMED THAT THE LOT WAS PRODUCED IN COMPLIANCE WITH ALL REQUIREMENTS, WITH NO DEVIATIONS OR MAINTENANCE EVENTS NOTED. SAMPLES WERE REQUESTED; HOWEVER, THE CUSTOMER CONFIRMED THAT NO SAMPLES WERE AVAILABLE FOR ANALYSIS. CONSEQUENTLY, AN INVESTIGATION WAS CONDUCTED USING REFERENCE SAMPLES, AND NO FAILURES RELATED TO THE COMPLAINT WERE IDENTIFIED. NO PHOTO EVIDENCE WAS PROVIDED, SO THE ASSESSMENT WAS BASED ON DOCUMENTATION AND REFERENCE SAMPLE ANALYSIS ONLY. BASED ON THESE RESULTS, NO MANUFACTURING OR QUALITY ISSUES WERE IDENTIFIED. THE RECORD WILL BE CLOSED AND MONITORED THROUGH ROUTINE TRACKING AND TRENDING

Description of Event or Problem · 0

REFERENCE NUMBER (B)(4). EVENT OCCURRED IN THE UNITED STATES. IT WAS REPORTED THAT THE PATIENT FACED HYPERGLYCEMIA ON (B)(6) 2025 DUE TO KINKED CANNULA. THE BLOOD GLUCOSE LEVEL WAS 624MG/DL AT THE TIME OF EVENT AND THE PATIENT GOT TREATED WITH INTRAVENOUS FLUIDS OF SALINE AND INSULIN. THE PATIENT WAS RELEASED FROM THE HOSPITAL ON (B)(6) 2025. NO FURTHER INFORMATION AVAILABLE.

Description of Event or Problem · 0

TO DATE NO ADDITIONAL PATIENT OR EVENT DETAILS HAVE BEEN RECEIVED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
114945 INSET UNO INSET I 60/6 GREY TCAP 10PK INT FPA UNOMEDICAL DEVICES S.A. DE C.V. 1001680 6013375 05705244016620

Patients

Seq Age Sex Outcome Treatment
1 64 YR Female Required Intervention| H