FDA Adverse Event Malfunction Summary report: N

0

MDR report key: 24051476 · Received January 13, 2026

Report

Report Number
3003442380-2025-19043
Event Type
Malfunction
Date Received
January 13, 2026
Date of Event
December 21, 2025
Report Date
January 14, 2026
Manufacturer
UNOMEDICAL A/S
Product Code
FPA
UDI-DI
05705244016651
PMA / PMN Number
K032854
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

ADDITIONAL INFORMATION - THIS MDR IS BEING SUBMITTED TO INCLUDE THE BELOW: H6: INVESTIGATION RESULTS UNDER TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, INVESTIGATION CONCLUSIONS. H11: INVESTIGATION SUMMARY. COMPLAINT INVESTIGATION RESULTS: ELECTRONIC QUALITY MANAGEMENT SYSTEM (EQMS) SEARCH A QUERY WAS RUN IN THE EQMS ON 14-JAN-2026 AGAINST "LOT NUMBER 6012832 AND SIMILAR MALFUNCTION CODES OCCLUSION IN THE TUBING AND/OR TUBING CONNECTORS- REDUCED FLOW, OCCLUSION IN THE TUBING AND/OR TUBING CONNECTORS (E.G., OCCLUSION ALARM FROM THE INFUSION PUMP MAY HAVE SOUNDED), OCCLUSION IN THE TUBING AND/OR TUBING CONNECTORS- BLOCKAGE. THE REVIEW CONFIRMED THAT LOT 6012832 AND THE IDENTIFIED FAILURE MODE ARE NOT ASSOCIATED WITH ANY NCRS OR CORRECTIVE AND PREVENTIVE ACTION (CAPA)S OF THE SAME OR SIMILAR NATURE. SIMILAR COMPLAINTS SEARCH: A QUERY WAS RUN IN THE EQMS ON 14-JAN-2026 AGAINST "LOT NUMBER" CRITERIA EQUAL 6012832 AND SIMILAR MALFUNCTION CODES OCCLUSION IN THE TUBING AND/OR TUBING CONNECTORS-REDUCED FLOW, OCCLUSION IN THE TUBING AND/OR TUBING CONNECTORS- BLOCKAGE), OCCLUSION IN THE TUBING AND/OR TUBING CONNECTORS (E.G.,OCCLUSION ALARM FROM THE INFUSION PUMP MAY HAVE SOUNDED). THE COUNT OF COMPLAINTS IS 1. THE COMPLAINT NUMBER IS 2498227. DEVICE HISTORY RECORD (DHR) REVIEW THE LOT 6012832 WAS MANUFACTURED ACCORDING TO THE WORK INSTRUCTION (WI) VERSION 122 AND MANUFACTURED IN THE LINE 08 ON 08-APR-2025 WITH A TOTAL OF (B)(4) UNITS. THE SUB-ASSEMBLY, GLUING OF TUBING THE LOT 5C06125 WAS MANUFACTURED ACCORDING TO THE WI VERSION 03 AND MANUFACTURED IN THE MACHINE ITL03, ON 21-APR-2025, WITH A TOTAL OF (B)(4) UNITS. THE DEVICE HISTORY RECORD (DHR) WAS REVIEWED IN ACCORDANCE WITH APPLICABLE PROCEDURES. ALL REQUIRED IN-PROCESS AND FINAL TESTS WERE COMPLETED AND MET SPECIFIED REQUIREMENTS. NO DEVIATIONS WERE IDENTIFIED, AND NO MAINTENANCE EVENTS WERE RECORDED THAT RELATE TO THE COMPLAINT CODE. CONCLUSION: DHR REVIEW SUPPORTS COMPLIANCE WITH MANUFACTURING AND QUALITY REQUIREMENTS; NO ISSUES NOTED. VISUAL EVIDENCE REVIEW: NO PHOTO WAS PROVIDED TO SUPPORT VISUAL CONFIRMATION OF THE REPORTED ISSUE. CONCLUSION: UNABLE TO PERFORM VISUAL VERIFICATION; ASSESSMENT BASED ON AVAILABLE DOCUMENTATION ONLY. CAPA DETERMINATION ASSESSMENT - CONCLUSION SUMMARY OF COMPLAINT INVESTIGATION: BASED ON THE INVESTIGATION, NO FURTHER INVESTIGATION IS REQUIRED. THE RECORD WILL BE CLOSED AND MONITORED THROUGH TRACKING AND TRENDING PER WI (MONTHLY TRIPS AND ALERTS). CONCLUSION SUMMARY OF COMPLAINT INVESTIGATION: AS A RESULT OF THE FOLLOWING A COMPREHENSIVE REVIEW WAS CONDUCTED, INCLUDING EQMS QUERIES, SIMILAR COMPLAINT SEARCHES, DEVICE HISTORY RECORD REVIEW, VISUAL EVIDENCE ASSESSMENT, AND CAPA DETERMINATION. NO NCRS OR CAPAS OF THE SAME OR SIMILAR NATURE WERE FOUND FOR LOT 6012832 AND RELATED MALFUNCTION CODES FOR OCCLUSION IN THE TUBING AND/OR TUBING CONNECTORS (E.G. OCCLUSION ALARM FROM THE INFUSION PUMP MAY HAVE SOUNDED). ONE MORE COMPLAINT WAS IDENTIFIED FOR THIS LOT, BUT NO TREND OR SYSTEMIC ISSUE WAS DETECTED. THE MANUFACTURING RECORDS CONFIRMED THAT THE LOT WAS PRODUCED IN COMPLIANCE WITH ALL REQUIREMENTS, WITH NO DEVIATIONS OR MAINTENANCE EVENTS NOTED. NO PHOTO EVIDENCE WAS PROVIDED, SO THE ASSESSMENT WAS BASED ON DOCUMENTATION ONLY. BASED ON THESE RESULTS, NO MANUFACTURING OR QUALITY ISSUES WERE IDENTIFIED, AND NO FURTHER INVESTIGATION IS REQUIRED. THE RECORD WILL BE CLOSED AND MONITORED THROUGH ROUTINE TRACKING AND TRENDING.

Description of Event or Problem · 0

TO DATE NO ADDITIONAL PATIENT OR EVENT DETAILS HAVE BEEN RECEIVED.

Description of Event or Problem · 0

REFERENCE NUMBER: (B)(4). EVENT OCCURRED IN THE UNITED STATES. IT WAS REPORTED THAT THE PATIENT FACED INSULIN FLOW BLOCKED ALARM EVENT ON (B)(6) 2025. THE BLOCKAGE WAS IN THE TUBING. THE PATIENT'S BLOOD GLUCOSE WAS HIGH AND WAS TREATED WITH BOLUS. NO FURTHER INFORMATION AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
115283 0 UNO INSET I 60/9 GREY TCAP 10PK INT FPA UNOMEDICAL A/S 1001681 6012832 05705244016651

Patients

Seq Age Sex Outcome Treatment
1 60 YR Female