FDA Adverse Event Malfunction Summary report: N

PROWLER SELECT

MDR report key: 24047569 · Received January 12, 2026

Report

Report Number
3008114965-2026-00085
Event Type
Malfunction
Date Received
January 12, 2026
Date of Event
December 23, 2025
Report Date
January 12, 2026
Manufacturer
MEDOS INTERNATIONAL SARL
Product Code
QJP
UDI-DI
10886704028888
PMA / PMN Number
K210838
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CH
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

MANUFACTURER¿S REF. NO: (B)(4). INFORMATION REGARDING PATIENT IDENTIFIER, DATE OF BIRTH, AGE, SEX, GENDER, WEIGHT, RACE, AND ETHNICITY WERE NOT PROVIDED. SECTION E.1: THE INITIAL REPORTER PHONE: (B)(6). BASED ON COMPLAINT INFORMATION, THE DEVICE IS NOT AVAILABLE TO BE RETURNED FOR ANALYSIS. A REVIEW OF MANUFACTURING DOCUMENTATION ASSOCIATED WITH THIS LOT: (31651253) PRESENTED NO ISSUES DURING THE MANUFACTURING OR INSPECTION PROCESSES RELATED TO THE REPORTED COMPLAINT. THERE WERE NO INTERNAL ACTIONS RELATED TO DEVICE MANUFACTURE OR INSPECTION. PRODUCT ANALYSIS CANNOT BE CONDUCTED AS THE PRODUCT WAS NOT RETURNED FOR ANALYSIS. NO DETERMINATION OF CAUSES AND POSSIBLE CONTRIBUTING FACTORS COULD BE MADE. AS SUCH, THE INVESTIGATION WILL BE CLOSED. WITH THE LIMITED INFORMATION AVAILABLE AND WITHOUT THE PRODUCT AVAILABLE FOR ANALYSIS, THE REPORTED ISSUE DOCUMENTED IN THE COMPLAINT COULD NOT BE CONFIRMED. BASED ON THE MANUFACTURING DOCUMENTATION REVIEW, THERE IS NO INDICATION THAT THE EVENT IS RELATED TO THE DEVICE MANUFACTURING PROCESS. THE EXACT CAUSE OF THE EVENT COULD NOT BE CONCLUSIVELY DETERMINED; HOWEVER, IT IS POSSIBLE THAT CIRCUMSTANCES OF THE PROCEDURE AND / OR DEVICE MANIPULATION / INTERACTION MAY HAVE CONTRIBUTED TO THE REPORTED FAILURE. AS PART OF THE POST MARKET SURVEILLANCE PROGRAM, INFORMATION FROM THIS COMPLAINT IS TRENDED FOR STATISTICAL SIGNALS AND CORRECTIVE / PREVENTIVE ACTION MAY BE TRIGGERED LATER. SINCE THERE WAS NO EVIDENCE TO SUGGEST THE EVENT WAS RELATED TO A MANUFACTURING OR DESIGN ISSUE, NO CORRECTIVE ACTIONS WILL BE TAKEN AT THIS TIME. THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY CERENOVUS, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, CERENOVUS, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL REPORT, A FOLLOW-UP REPORT WILL BE FILED AS APPROPRIATE. THE MANUFACTURER WILL SUBMIT A SUPPLEMENTAL REPORT IF NEW FACTS ARISE WHICH MATERIALLY ALTER INFORMATION SUBMITTED IN A PREVIOUS MDR REPORT. ADDITIONAL INFORMATION WILL BE SUBMITTED WITHIN 30 DAYS OF RECEIPT.

Description of Event or Problem · 0

THE HEALTHCARE PROFESSIONAL REPORTED THAT DURING AN ENDOVASCULAR EMBOLIZATION PROCEDURE, THE 4.5MM X 22MM NO DISTAL TIP ENTERPRISE VASCULAR RECONSTRUCTION DEVICE (ENC452200 / 9926000) WAS IMPEDED IN THE HUB OF THE CONCOMITANT 150CM X 5CM PROWLER SELECT PLUS MICROCATHETER (606S255X / 31651253) AND COULD NOT BE INSERTED INTO THE MICROCATHETER. THE PHYSICIAN REMOVED THE MICROCATHETER AND THE STENT FROM THE PATIENT AND DELIVERED THE ORIGINAL STENT INTO THE MICROCATHETER AGAIN, BUT THE SAME ISSUE WAS ENCOUNTERED. THE PHYSICIAN REMOVED THE DEVICES AND REPLACED THEM WITH DEVICES FROM OTHER MANUFACTURERS TO COMPLETE THE PROCEDURE. THERE WAS NO REPORT OF ANY NEGATIVE PATIENT IMPACT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
110388 PROWLER SELECT CATHETER, PERCUTANEOUS, NEUROVASCULATURE QJP MEDOS INTERNATIONAL SARL 31651253 10886704028888

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown