FDA Adverse Event Malfunction Summary report: N

ATELLICA SH PRIME

MDR report key: 24044111 · Received January 12, 2026

Report

Report Number
2432235-2026-00009
Event Type
Malfunction
Date Received
January 12, 2026
Date of Event
December 24, 2025
Report Date
January 12, 2026
Manufacturer
SIEMENS HEALTHCARE DIAGNOSTICS INC.
Product Code
JJE
UDI-DI
630414002439
PMA / PMN Number
K161954
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IT
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

A DAMAGED SAMPLE ID (SID) BARCODE LABEL WAS MISREAD WHEN LOADED AND SCANNED BY THE ATELLICA SAMPLE HANDLER PRIME (SHP). A SIEMENS REGIONAL SUPPORT SPECIALIST (RSC) INVESTIGATED THE REPORTED ISSUE AND FOUND THAT THE SID BARCODE LABEL WAS DAMAGED / TORN. THE SID LABEL WAS TORN WHEN APPLIED TO THE TUBE AND WAS NOT DAMAGED BY THE ATELLICA SHP. THE SHP WAS CONFIGURED TO ALLOW FOR A BARCODE SYMBOLOGY THAT DOES NOT CONTAIN A CHECK DIGIT. A CHECK DIGIT IS USED TO PREVENT MISREADS. THE SID BARCODE FORMAT IS UNKNOWN. THE CAUSE OF THE MISREAD SID IS DUE TO A USE ERROR OF APPLYING A DAMAGED BARCODE LABEL THAT WAS PRINTED WITH A BARCODE SYMBOLOGY THAT DOES NOT INCLUDE A CHECK DIGIT. THE ANALYZER IS PERFORMING WITHIN SPECIFICATIONS. NO FURTHER EVALUATION OF THIS ANALYZER IS NEEDED.

Description of Event or Problem · 0

A DAMAGED SAMPLE ID (SID) BARCODE LABEL WAS MISREAD WHEN LOADED ON AND SCANNED BY THE ATELLICA SAMPLE HANDLER PRIME (SHP). TESTS WERE PERFORMED ON THE MISREAD SID. THE TEST RESULTS WERE NOT RELEASED TO A PHYSICIAN. THE SID BARCODE LABEL WAS REPRINTED, AND THE SAME TUBE WAS RELABELED AND THEN REPROCESSED ON THE SHP. THERE ARE NO KNOWN REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
103084 ATELLICA SH PRIME ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE JJE SIEMENS HEALTHCARE DIAGNOSTICS INC. 11069001 630414002439

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown