ALINITY I STAT HIGH SENSITIVE TROPONIN-I REAGENT KIT
Report
- Report Number
- 3005094123-2026-00015
- Event Type
- Malfunction
- Date Received
- January 12, 2026
- Date of Event
- December 30, 2025
- Report Date
- March 23, 2026
- Manufacturer
- ABBOTT IRELAND DIAGNOSTICS DIVISION
- Product Code
- MMI
- UDI-DI
- 00380740186029
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AS
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
A COMPLAINT INVESTIGATION WAS CONDUCTED FOR THE ALINITY I STAT HIGH SENSITIVE TROPONIN-I REAGENT, LOT 78014UD00. DATA AND INFORMATION PROVIDED BY THE CUSTOMER WERE REVIEWED AND SUPPORT THE COMPLAINT ISSUE. REVIEW OF TRACKING AND TRENDING BY LIST NUMBER AND BY COMPLAINT LOT DID NOT IDENTIFY ANY TRENDS. DEVICE HISTORY REVIEW DID NOT IDENTIFY ISSUES ASSOCIATED WITH THE CUSTOMER¿S OBSERVATION. THE OVERALL PERFORMANCE OF ALINITY I STAT HIGH SENSITIVE TROPONIN-I REAGENTS IN THE FIELD WAS REVIEWED USING DATA FROM CUSTOMERS WORLDWIDE. THE PATIENT MEDIAN VALUES FOR LOTS 78011UD00 AND 78013UD00 (WHICH CONTAIN THE SAME BULK MATERIAL AS LOT 78014UD00) ARE WITHIN THE ESTABLISHED LIMITS AND COMPARABLE TO THE HISTORICAL REAGENT LOT PERFORMANCE. REVIEW OF APPLICABLE FIELD DATA SUGGESTS THAT THE PERFORMANCE IS ACCEPTABLE. LABELING WAS REVIEWED WHICH ADEQUATELY ADDRESSES THE CURRENT ISSUE. BASED ON OUR INVESTIGATION, NO SYSTEMIC ISSUE OR DEFICIENCY WAS IDENTIFIED WITH THE ALINITY I STAT HIGH SENSITIVE TROPONIN-I REAGENT, LOT 78014UD00.
AN EVALUATION IS IN PROCESS. A FOLLOW UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE. ALL AVAILABLE PATIENT INFORMATION WAS INCLUDED. ADDITIONAL PATIENT DETAILS ARE NOT AVAILABLE. SECTION A1 PATIENT IDENTIFIER COMPLETE SID: (B)(6) ( ALL SIDS WERE FROM THE SAME PATIENT).
THE CUSTOMER REPORTED FALSELY ELEVATED ALINITY I STAT HIGH SENSITIVE TROPONIN-I RESULT ON A 42-YEAR-OLD MALE PATIENT. ALL SID LISTED BELOW ARE FROM THE SAME PATIENT. RESULTS WERE REPORTED TO MEDICAL PROVIDER. THE FOLLOWING RESULTS WERE PROVIDED: (B)(6) 2025, SID (B)(6), INITIAL RESULT = 223 NG/L, REPEAT RESULT ON ANOTHER PROCESSING MODULE (B)(6) = < 2 NG/L. (B)(6) 2025, SID (B)(6), RESULTS WERE = 4 NG/L & REPEAT RESULT ON ANOTHER PROCESSING MODULE (B)(6) = 3 NG/L. (B)(6) 2025, SID (B)(6) = RESULTS WERE = < 2 NG/L, (B)(6) 2025 = < 2 NG/L & < 2 NG/L AND (B)(6) 2025, RESULT ON ANOTHER PROCESSING MODULE (B)(6) = < 2 NG/L. THE CUSTOMER USES REFERENCE RANGE < 26 NG/L. THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.
THE CUSTOMER REPORTED FALSELY ELEVATED ALINITY I STAT HIGH SENSITIVE TROPONIN-I RESULT ON A 42-YEAR-OLD MALE PATIENT. ALL SID LISTED BELOW ARE FROM THE SAME PATIENT. RESULTS WERE REPORTED TO MEDICAL PROVIDER. THE FOLLOWING RESULTS WERE PROVIDED: ON (B)(6) 2025, SID (B)(6), INITIAL RESULT = 223 NG/L, REPEAT RESULT ON ANOTHER PROCESSING MODULE (B)(6) = < 2 NG/L. ON (B)(6) 2025, SID (B)(6), RESULTS WERE = 4 NG/L & REPEAT RESULT ON ANOTHER PROCESSING MODULE (B)(6) = 3 NG/L. ON (B)(6) 2025, SID (B)(6) = RESULTS WERE = < 2 NG/L, (B)(6) 2025 = < 2 NG/L & < 2 NG/L AND (B)(6) 2025, RESULT ON ANOTHER PROCESSING MODULE (B)(6) = < 2 NG/L. THE CUSTOMER USES REFERENCE RANGE < 26 NG/L. THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 277888 | ALINITY I STAT HIGH SENSITIVE TROPONIN-I REAGENT KIT | IMMUNOASSAY METHOD, TROPONIN SUBUNIT | MMI | ABBOTT IRELAND DIAGNOSTICS DIVISION | 78014UD00 | 00380740186029 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 42 YR | Male | ALNTY I PROCESSING MODU, 03R65-01, (B)(6).| ALNTY I PROCESSING MODU, 03R65-01, (B)(6). |