FDA Adverse Event Death Summary report: N

BERMAN ANGIOGRAPHIC CATHETER

MDR report key: 24019181 · Received January 9, 2026

Report

Report Number
MW5181976
Event Type
Death
Date Received
January 9, 2026
Date of Event
October 13, 2025
Report Date
January 5, 2026
Manufacturer
ARROW INTERNATIONAL LLC (SUBSIDIARY OF TELEFLEX INCORPORATED)
Product Code
DYG
Adverse Event
Yes
Report Source
Voluntary report
Reporter Occupation
NURSE
Health Professional
Yes

Narratives

Description of Event or Problem · 0

A BERMAN ANGIOGRAPHIC CATHETER WAS INADVERTENTLY USED INSTEAD OF THE INTENDED BALLOON WEDGE CATHETER, RESULTING IN A RUPTURED PULMONARY ARTERY DURING AN OUTPATIENT PROCEDURE. PATIENT REQUIRED MULTIPLE INTERVENTIONS TO SUSTAIN LIFE & ICU ADMISSION, HOWEVER ULTIMATELY EXPIRED. BOTH CATHETERS HAVE SIMILAR PACKAGING AND APPEARANCE & ARE MANUFACTURED BY SAME COMPANY (ARROW/TELEFLEX). CONCERN FOR RISK OF RECURRENCE AT OTHER INSTITUTIONS DUE TO SIMILARITY IN PACKAGING & DEVICE APPEARANCE. REF REPORT: MW5181977. PT CODE: 2152. DEVICE CODES: 4081, 2958, 4001.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
80637 BERMAN ANGIOGRAPHIC CATHETER CATHETER, FLOW DIRECTED DYG ARROW INTERNATIONAL LLC (SUBSIDIARY OF TELEFLEX INCORPORATED) AI-07137

Patients

Seq Age Sex Outcome Treatment
1 72 YR Female Death| R| H