FDA Adverse Event Malfunction Summary report: N

MENTOR MEMORYGEL BREAST IMPLANT

MDR report key: 24013980 · Received January 8, 2026

Report

Report Number
1645337-2026-00333
Event Type
Malfunction
Date Received
January 8, 2026
Date of Event
December 18, 2025
Manufacturer
MENTOR TEXAS
Product Code
FTR
PMA / PMN Number
P030053
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

SINCE THE DEVICE HAS NOT BEEN RETURNED FOR ANALYSIS, NO PRODUCT FAILURE ANALYSIS CAN BE CONDUCTED, AND NO DETERMINATION OF POSSIBLE CONTRIBUTING FACTORS CAN BE MADE. AS SUCH, THE INVESTIGATION WILL BE CLOSED. IF THE COMPLAINT DEVICE IS RECEIVED IN THE FUTURE, THE INVESTIGATION WILL BE REOPENED AND CONDUCTED AS APPROPRIATE. SINCE NO LOT NUMBER WAS PROVIDED, NO MANUFACTURING RECORD EVALUATION COULD BE PERFORMED. REASON FOR DEVICE EXPLANT AND/OR REOPERATION: RUPTURE. MENTOR IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH MENTOR HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, MENTOR, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, MENTOR, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF CERTAIN INFORMATION IS UNKNOWN, NOT AVAILABLE OR DOES NOT APPLY, THE SECTION/FIELD OF THE FORM IS LEFT BLANK. D4: UDI: AS THE LOT NUMBER FOR THE DEVICE INVOLVED IN THE EVENT WAS NOT PROVIDED, THE FULL UDI IS CURRENTLY NOT AVAILABLE. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).

Additional Manufacturer Narrative · 0

SUSPECT DEVICE RECEIVED ON FEBRUARY 04, 2026. DEVICE EVALUATION COMPLETED ON FEBRUARY 11, 2026: THE PRODUCT WAS RETURNED TO MENTOR FOR EVALUATION. MENTOR CONDUCTED A VISUAL INSPECTION, MICROSCOPIC EXAMINATION, AND THICKNESS MEASUREMENT OF THE RETURNED DEVICE. THE SUSPECT DEVICE IDENTITY REVEALED AS LOT: 2112606, CAT:3505295BC. VISUAL ANALYSIS OF THE RETURNED DEVICE REVEALED THAT THE BREAST IMPLANT WAS FOUND TO HAVE A TEAR ON THE POSTERIOR VIEW, MEASURING APPROXIMATELY 1.5 CM. MICROSCOPIC EXAMINATION WAS PERFORMED, AND THE CAUSE OF THE TEAR COULD NOT BE IDENTIFIED. THEREFORE, A THICKNESS MEASUREMENT WAS CONDUCTED ON THE SHELL AT THE TEAR, AND THE THICKNESS WAS WITHIN MANUFACTURING SPECIFICATIONS. MENTOR ALSO PERFORMED A MANUFACTURING RECORD EVALUATION RELATED TO THE REPORTED COMPLAINT FOR THE FINISHED DEVICE LOT NUMBER. NO MANUFACTURING NON-CONFORMANCES WERE IDENTIFIED AS PART OF THIS EVALUATION. ALTHOUGH NO CONCLUSION COULD BE REACHED ON THE CAUSE OF THE REPORTED EVENT, THE INSTRUCTIONS FOR USE CONTAIN THE FOLLOWING CAUTION: IT IS POSSIBLE TO DAMAGE THE IMPLANT DURING INSERTION. ENSURE THAT EXCESSIVE FORCE IS NOT APPLIED TO A VERY SMALL AREA OF THE SHELL DURING INSERTION OF THE DEVICE THROUGH THE INCISION. INSTEAD, APPLY FORCE OVER AS LARGE AN AREA OF THE IMPLANT AS POSSIBLE. AVOID PUSHING THE DEVICE INTO PLACE WITH ONE OR TWO FINGERS IN A LOCALIZED AREA, AS THIS MAY CREATE AN AREA OF WEAKNESS ON THE SHELL. IN ADDITION, AN INCISION SHOULD BE OF APPROPRIATE LENGTH TO ACCOMMODATE THE STYLE, SIZE, AND PROFILE OF THE IMPLANT. THIS WILL REDUCE THE POTENTIAL FOR CREATING EXCESSIVE STRESS ON THE IMPLANT DURING INSERTION AND WILL AVOID THE RISK OF DAMAGE TO THE IMPLANT. AS PART OF MENTOR'S QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED ACCORDING TO APPROVED SPECIFICATIONS. BASED ON THE RESULTS OF THIS INVESTIGATION, IT WAS DETERMINED THAT THE PRODUCT MET THE MANUFACTURING RELEASE CRITERIA: NO CORRECTIVE AND PREVENTIVE ACTION (CAPA) IS REQUIRED NOW. THE MANUFACTURING RECORD EVALUATION (MRE) WAS REVIEWED, AND NO ANOMALIES WERE FOUND RELATED TO THIS COMPLAINT. IN ADDITION, THE MRE REVIEW VERIFIES THAT THE DEVICE WAS MANUFACTURED IN ACCORDANCE WITH DOCUMENTED SPECIFICATION AND PROCEDURES. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).

Description of Event or Problem · 0

A FEMALE PATIENT WHO UNDERWENT AN UNKNOWN BREAST SURGERY WITH AN UNKNOWN MENTOR SILICONE PROSTHESIS EXPERIENCED UNKNOWN SIDED RUPTURE INTRAOPERATION. THE IMPLANT RUPTURED WHILE THE SURGEON WAS PLACING IT. NO ADVERSE EVENT REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
65726 MENTOR MEMORYGEL BREAST IMPLANT PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED FTR MENTOR TEXAS 2112606

Patients

Seq Age Sex Outcome Treatment
1 NA Female