IMMAGE® RHEUMATOID FACTOR REAGENT
Report
- Report Number
- 2050012-2012-00037
- Event Type
- Malfunction
- Date Received
- January 6, 2012
- Date of Event
- November 29, 2011
- Report Date
- December 6, 2011
- Manufacturer
- BECKMAN COULTER, INC.
- Product Code
- DHR
- PMA / PMN Number
- K963048
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER
Narratives
NO SAMPLE INFORMATION WAS PROVIDED. THE CUSTOMER CLAIMED QC WAS WITHIN THE ESTABLISHED RANGES. THE CUSTOMER WAS SENT ANOTHER LOT OF REAGENT (LOT M103174) FAILURE MODE APPEARS TO BE THE CALIBRATION OF THE REAGENT LOT, M012376. THE RELATED EVENTS AT THIS CUSTOMER SITE ARE REPORTED IN BELOW LISTED MDRS: 2050012-2012-00036, 2050012-2012-00038.
A CUSTOMER REPORTED TO BECKMAN COULTER INC. (BCI) THAT ERRONEOUSLY ELEVATED RHEUMATOID FACTOR (RF) RESULTS WERE GENERATED BY IMMAGE 800 IMMUNOCHEMISTRY SYSTEM ON (B)(6) 2011. IMMAGE RHEUMATOID FACTOR REAGENT LOT M012376 WAS USED FOR THE ASSAYS. THE ERRONEOUS RESULTS WERE NOT REPORTED OUT OF THE LABORATORY. REPEAT TESTING AFTER RECALIBRATION WITH THE SAME LOT OF RF REAGENT PRODUCED RESULTS WITHIN THE NORMAL REFERENCE RANGE, BELOW 20 IU/ML. REPEAT TESTING RESULTS WERE NOT PROVIDED. THERE WAS NO REPORT OF EFFECT TO PATIENT TREATMENT. THIS REPORT DOCUMENTS THE EVENT ON (B)(6) 2011 WHEN ONE (1) ERRONEOUSLY ELEVATED RESULT WAS GENERATED. IMMAGE 800 IMMUNOCHEMISTRY SYSTEM: CATALOGUE NUMBER - A15445, SERIAL NUMBER - (B)(4), DATE OF MANUFACTURE - 08/18/2005, PRODUCT CODE - JQX, NEPHELOMETER, FOR CLINICAL USE, 510(K) NUMBER - K962294.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | IMMAGE® RHEUMATOID FACTOR REAGENT | RHEUMATOID FACTOR TEST REAGENT | DHR | BECKMAN COULTER, INC. | IMMAGE RF | M012376 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |