CRYO SPRAY ABLATION

MDR report key: 2400641 · Received January 3, 2012

Report

Report Number
3004534508-2011-00006
Event Type
Injury
Date Received
January 3, 2012
Date of Event
December 6, 2011
Report Date
January 3, 2012
Manufacturer
CSA MEDICAL
Product Code
GEH
PMA / PMN Number
K060555
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE CRYOSPRAY SYSTEM OPERATED NORMALLY AND MET ALL ACCEPTANCE CRITERIA DURING A SUBSEQUENT FIELD VERIFICATION TEST. NO ERRORS OF THE BUILD-IN TEST PERFORMED AT EACH SYSTEM POWER-UP WERE RECORDED. THE PHYSICIAN BELIEVES THE PERFORATION WAS RELATED TO EXPANDING GAS FROM THE CRYOSPRAY PROCEDURE AS A RESULT OF THE SMALLER NG VENT TUBE USED IN PLACE OF THE RECOMMENDED CRYOGEN DECOMPRESSION TUBE SUPPLIED WITH THE CSA DISPOSABLE KIT AND SPECIFIED IN THE INSTRUCTIONS FOR USE.

Description of Event or Problem · 1

PT PRESENTED FOR CRYOSPRAY TREATMENT OF ESOPHAGEAL CANCER ON (B)(6) 2011. A KNOWN STRICTURE PREVENTED THE CSA SUPPLIED 20 FRENCH DUAL LUMEN CSA CRYOGEN DECOMPRESSION (VENT) FROM BEING PLACED IN THE PT AS SPECIFIED IN THE INSTRUCTIONS FOR USE (IFU). INSTEAD, A 14 FRENCH NASOGASTRIC (NG) TUBE WAS INSERTED IN THE PT AND USED FOR EVACUATING THE GAS. THE PT'S ABDOMEN WAS MONITORED FOR DISTENTION DURING THE PROCEDURE PER THE IFU. AFTER 10 SECONDS INTO THE FIRST FREEZE, THE ABDOMINAL MONITORING NURSE NOTED ACUTE DISTENTION. THE PHYSICIAN STOPPED THE CRYOSPRAY, ENSURED SUCTION WAS ON, AND WAITED FOR THE GAS TO DISSIPATE. THE ABDOMEN REMAINED DISTENDED LONGER THAN EXPECTED. THE NG TUBE WAS REMOVED AND THE PT'S ESOPHAGUS AND ABDOMEN WERE EXAMINED ENDOSCOPICALLY. THE PHYSICIAN OBSERVED A PERFORATION IN THE STOMACH. THE PT WAS TAKEN TO SURGERY APPROX 30 MINUTES AFTER THE PERFORATION WAS DISCOVERED. A LAPAROTOMY WAS PERFORMED TO SURGICALLY REPAIR THE PERFORATION. THE PT WAS DISCHARGED FROM THE HOSPITAL APPROXIMATELY 1 WEEK POST OPERATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CRYO SPRAY ABLATION CRYO SPRAY ABLATION GEH CSA MEDICAL CC2-NAM NA

Patients

Seq Age Sex Outcome Treatment
1 83 YR Hospitalization| R