FDA Adverse Event Malfunction Summary report: N

NIBP CUFFS

MDR report key: 23988391 · Received January 7, 2026

Report

Report Number
9610816-2026-100017
Event Type
Malfunction
Date Received
January 7, 2026
Date of Event
December 23, 2025
Report Date
February 20, 2026
Manufacturer
PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH
Product Code
OED
UDI-DI
00884838114784
PMA / PMN Number
K071885
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
LH
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

A PHILIPS PRODUCT INDUSTRIALIZATION ENGINEER (PIE) AND REGULATORY AFFAIRS SPECIALIST (RAS) REVIEWED THE AVAILABLE DATA AND CONFIRMED THAT THE WORD CUFF WAS MISSING FROM THE PRODUCT; THIS AFFECTED THE GRAPHIC PRINTING ON THE CUFF BODY FOR THIS ONE DEVICE. BASED ON THE RESULTS OF THE ANALYSIS, IT WAS DETERMINED THAT THE PRODUCT HAS MALFUNCTIONED AND THE MOST LIKELY CAUSE OF THE REPORTED PROBLEM WAS A PRINTING ISSUE ON THIS DEVICE. THE ISSUE WAS BEING RESOLVED BY A REVIEW ALL CUFFS' CDLMS (GENTLE LS-200080, EASY LS-200081, COMFORT LS-200078, NEONATAL LS-200079, AND AIR HOSES LS-200082) TO CHECK IF THERE ARE MARKINGS TO BE ADDED TO THE CDLM. THE DATA ENTERED IN THIS COMPLAINT RECORD WILL BE UTILIZED FOR PRODUCT QUALITY AND SAFETY IMPROVEMENTS PER THE POST MARKET SURVEILLANCE AND RISK MANAGEMENT PROCESSES. IF ADDITIONAL INFORMATION IS RECEIVED, THE COMPLAINT FILE WILL BE REOPENED FOR FURTHER INVESTIGATION.

Additional Manufacturer Narrative · 0

PHILIPS IS IN THE PROCESS OF OBTAINING ADDITIONAL INFORMATION CONCERNING THIS EVENT AND THE COMPLAINT IS STILL UNDER INVESTIGATION. A FINAL REPORT WILL BE SUBMITTED ONCE THE INVESTIGATION IS COMPLETE.

Description of Event or Problem · 0

PHILIPS RECEIVED A COMPLAINT ON AN EASY CARE CUFF 1 HOSE, LARGE ADULT XL INDICATING THAT THE CENTRAL COE MV (CENTER OF EXCELLENCE MARKET VERIFICATION) PERFORMED A VERIFICATION OF LABELING AND FOUND A LACK OF THE FULL NAME OF THE DEVICE ON THE PRODUCT, THE LABEL WORD "CUFF" MISSING, AND A LACK OF THE LEGAL MANUFACTURER'S ADDRESS AND INDICATOR. THIS WAS NOTED DURING INSPECTION; NO ADVERSE EVENT OR HARM WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
51463 NIBP CUFFS ANTIMICROBIAL BLOOD PRESSURE CUFF OED PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH EASY CARE CUFF,1 HOSE,LRG ADULT XL (1) 00884838114784

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown