FDA Adverse Event Death Summary report: N

FRESENIUS DIALYSIS DELIVERY SYSTEM

MDR report key: 2398365 · Received January 4, 2012

Report

Report Number
2937457-2011-00024
Event Type
Death
Date Received
January 4, 2012
Date of Event
October 15, 2011
Report Date
November 21, 2011
Manufacturer
FRESENIUS MEDICAL CARE NORTH AMERICA
Product Code
KDI
PMA / PMN Number
K994267
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
AL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A MACHINE INVESTIGATION BY THE REG EQUIPMENT SPECIALIST (RES) WAS CONDUCTED AND IT WAS DETERMINED THAT A UF VARIANCE OF APPROX 50 -60 ML/HR EXISTED. THE UF VARIANCE OF 240 ML WITH THE SET UF GOAL OF 3 KG OVER A 4 HOUR TREATMENT TIME IS NOT SIGNIFICANT TO ADVERSELY AFFECT THE PT. (B)(4).

Description of Event or Problem · 1

THE PT WAS HEMODIALYZING IN THE HOSP AND RECEIVING 4 UNITS OF PACKED RED BLOOD CELLS (RBCS) AS A RESULT OF GROSS RECTAL BLEEDING. HER BLOOD PRESSURE (BP) WAS VERY LOW (89/43) FROM THE BEGINNING AND HER DR, DR (B)(6) WAS AWARE. THE NURSE DID NOT SET THE HEMODIALYSIS MACHINE TO REMOVE EXCESS FLUID BECAUSE OF THE LOW BP, HOWEVER, THE MACHINE WAS SET TO REMOVE 150 ML/HR FOR THE 4 HOUR TREATMENT PERIOD. HER DIALYZER AND VENOUS BLOODLINES CLOTTED ABOUT HALF WAY THROUGH THE THIRD UNIT OF RBCS; SHE WAS NOT RECEIVING ANY HEPARIN. THE NURSE HAD RETURNED THE PT'S BLOOD SO THAT THE DIALYZER AND BLOOD LINES COULD BE REPLACED TO ALLOW THE PT TO RECEIVE THE REMAINING UNIT OF BLOOD AND COMPLETE HER TREATMENT. WHEN THE NURSE STARTED UNHOOKING THE PT FROM THE BLOODLINES, SHE NOTICED THE PT'S EYES ROLLING TO THE BACK OF HER HEAD. THE PT TOOK A REALLY DEEP BREATH, THE NURSE CALLED THE PT'S NAME SEVERAL TIMES BUT SHE DID NOT RESPOND. THE FLOOR NURSE WAS NOTIFIED AND A CODE WAS INITIATED. CARDIOPULMONARY RESUSCITATION (CPR) WAS STARTED UNTIL THE RESPONSE TEAM ARRIVED. THE PT WAS UNABLE TO BE RESUSCITATED AND SUBSEQUENTLY DIED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 FRESENIUS DIALYSIS DELIVERY SYSTEM HEMODIALYSIS MACHINE KDI FRESENIUS MEDICAL CARE NORTH AMERICA 2008K NA

Patients

Seq Age Sex Outcome Treatment
1 61 YR Death