FDA Adverse Event Malfunction Summary report: N

ELECSYS TROPONIN T HS

MDR report key: 23979022 · Received January 6, 2026

Report

Report Number
1823260-2026-00062
Event Type
Malfunction
Date Received
January 6, 2026
Date of Event
November 13, 2025
Report Date
February 23, 2026
Manufacturer
ROCHE DIAGNOSTICS
Product Code
MMI
PMA / PMN Number
K201441
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE QC WAS WITHIN RANGE. THERE WERE NO SAMPLE QUALITY-RELATED ALARMS. THE INVESTIGATION NOTED A GENERAL SUB-OPTIMAL PATIENT SAMPLE QUALITY AT THE CUSTOMER SITE. THE PROVIDED DATA DO NOT INDICATE AN ISSUE WITH THE ANALYZER OR THE REAGENT. THE INVESTIGATION DID NOT IDENTIFY A PRODUCT PROBLEM. THE CAUSE OF THE EVENT WAS CONSISTENT WITH A PREANALYTIC (SUB-OPTIMAL PATIENT SAMPLE QUALITY) ISSUE AT THE CUSTOMER SITE. PREANALYTICS ARE WITHIN THE CUSTOMER'S RESPONSIBILITY.

Additional Manufacturer Narrative · 0

THE COBAS E 801 ANALYTICAL UNIT SERIAL NUMBER WAS (B)(6). THE INVESTIGATION IS ONGOING.

Description of Event or Problem · 0

WE RECEIVED AN ALLEGATION ABOUT DISCREPANT RESULTS FOR 8 PATIENTS' SAMPLES TESTED WITH ELECSYS TROPONIN T HS ASSAY ON A COBAS E 801 ANALYTICAL UNIT. SAMPLE 1: INITIAL RESULT: 166 NG/L. REPEAT RESULT: 64.0 NG/L. ON (B)(6) 2025: SAMPLE 2: INITIAL RESULT: 25.7 NG/L. 1ST REPEAT RESULT: 20.6 NG/L. 2ND REPEAT RESULT: 20.7 NG/L. ON (B)(6) 2025: SAMPLE 3: INITIAL RESULT: 36.6 NG/L. 1ST REPEAT RESULT: 25.2 NG/L. 2ND REPEAT RESULT: 25.6 NG/L. SAMPLE 4: INITIAL RESULT: 73.1 NG/L. REPEAT RESULT: 60.8 NG/L. ON (B)(6) 2025: SAMPLE 5: INITIAL RESULT: 17.5 NG/L. 1ST REPEAT RESULT: 10.6 NG/L. 2ND REPEAT RESULT: 8.78 NG/L. ON (B)(6) 2025: SAMPLE 6: INITIAL RESULT: 16.4 NG/L. 1ST REPEAT RESULT: 12.6 NG/L. 2ND REPEAT RESULT: 11.9 NG/L. ON (B)(6) 2025: SAMPLE 7: INITIAL RESULT: 16.9 NG/L. REPEAT RESULT: 13.8 NG/L. SAMPLE 8: INITIAL RESULT: 36.9 NG/L. REPEAT RESULT: 46.9 NG/L.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
33512 ELECSYS TROPONIN T HS IMMUNOASSAY METHOD, TROPONIN SUBUNIT MMI ROCHE DIAGNOSTICS 84737601

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown