FDA Adverse Event
Injury
Summary report: N
SIGN I.M. NAIL
MDR report key: 2397175
·
Received December 27, 2011
Report
- Report Number
- 3034525-2011-00043
- Event Type
- Injury
- Date Received
- December 27, 2011
- Date of Event
- December 12, 2011
- Report Date
- December 27, 2011
- Manufacturer
- SIGN FRACTURE CARE INTERNATIONAL
- Product Code
- HSB
- PMA / PMN Number
- K022632
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PHYSICIAN
Narratives
Description of Event or Problem · 1
REPORT OF BROKEN IM NAIL IN DATABASE. SURGEON INDICATES HIGH FEMUR F/X REPAIRED, THEN PT COMPLAINED AND IT WAS FOUND THAT THE NAIL WAS BROKEN. INDICATIONS OF NON-UNION HIGH ON THE FEMUR ON AN OBESE PT W/FWB IS CAUSE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SIGN I.M. NAIL | SIGN I.M. NAIL | HSB | SIGN FRACTURE CARE INTERNATIONAL | UNK | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 34 YR |