EPOC BGEM TEST CARD
Report
- Report Number
- 3004959793-2011-00005
- Event Type
- Malfunction
- Date Received
- November 28, 2011
- Date of Event
- October 27, 2011
- Report Date
- November 28, 2011
- Manufacturer
- EPOCAL INC.
- Product Code
- CHL
- PMA / PMN Number
- K090109
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AL, US
- Reporter Occupation
- OTHER
Narratives
THE SINGLE USE TEST CARD IN QUESTION WAS NOT RETURNED. THE ACTUAL TEST DATA FROM THE EPOC DEVICE WAS MADE AVAILABLE. REPRESENTATIVE PRODUCT SAMPLES WERE REQUESTED, BUT HAVE NOT YET BEEN RETURNED FOR EVAL. BASED UPON THE AVAILABLE INFO, THE DISCREPANCY OBSERVED BY THE CUSTOMER WAS AN ISOLATED EVENT THAT OCCURRED ON TWO DIFFERENT BLOOD SPECIMENS COLLECTED IN DIFFERENT COLLECTION DEVICES. AN ASSIGNABLE CAUSE HAS NOT YET BEEN DETERMINED.
(B)(6) HAD A DISCREPANCY IN GLUCOSE VALUES BETWEEN THE EPOC AND KODAK ANALYZER. ON (B)(6) 2011 AT 13:02, THE EPOC REPORTED A GLUCOSE VALUE OF 361 ON A (B)(6) OLD DIABETIC PT. "THE PT WAS TREATED ON THE EPOC RESULTS, HOWEVER, MAINTENANCE OF PT WOULD NOT HAVE BEEN ANY DIFFERENT REGARDLESS. WE GOT THE MAIN LAB RESULTS BACK BEFORE THE INSULIN DRIP WAS STARTED. THERE WAS NO INJURY TO THE PT." A GLUCOSE VALUE OF 946 WAS GIVEN ON THE KODAK FUSION AT 13:43 ON (B)(6) 2011. THE SAME SAMPLE WAS NOT TESTED ON BOTH THE EPOC AND KODAK BUT THE SAMPLES WERE DRAWN AT THE SAME TIME. THE KODAK SAMPLE WAS DRAWN AT 13:08 USING A NEW SYRINGE AND WAS A RIGHT ANTECUBITAL VENOUS SAMPLE THAT WAS PUT INTO A LAB TUBE. THE EPOC SAMPLE WAS DRAWN FIRST. IT WAS A RIGHT ANTECUBITAL VENOUS SAMPLE DRAWN USING A PLAIN 3 CC SYRINGE WITHOUT ANTICOAGULANT. THERE WERE NO PROCEDURES BEING PERFORMED AND THE IV HADN'T BEEN STARTED WHEN THE SAMPLE WAS DRAWN. THE PT'S MOM GAVE THE PT 20 UNITS INSULIN HOME ABOUT 12:15 OR 12:30 ON (B)(6) 2011.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | EPOC BGEM TEST CARD | CHL, CEM, JFP, JPI, JGS, CGA, KHP | CHL | EPOCAL INC. | EPOC BGEM | 07-11222-00 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 16 YR | Other |