FDA Adverse Event Injury Summary report: N

MENTOR MEMORYGEL BREAST IMPLANT

MDR report key: 23959457 · Received January 5, 2026

Report

Report Number
1645337-2026-00115
Event Type
Injury
Date Received
January 5, 2026
Date of Event
October 1, 2025
Manufacturer
MENTOR TEXAS
Product Code
FTR
UDI-DI
81317000310
PMA / PMN Number
P030053
Removal / Correction Number
N/A
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MD, US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

SINCE THE DEVICE HAS NOT BEEN RETURNED FOR ANALYSIS, NO PRODUCT FAILURE ANALYSIS CAN BE CONDUCTED, AND NO DETERMINATION OF POSSIBLE CONTRIBUTING FACTORS CAN BE MADE. AS SUCH, THE INVESTIGATION WILL BE CLOSED. IF THE COMPLAINT DEVICE IS RECEIVED IN THE FUTURE, THE INVESTIGATION WILL BE REOPENED AND CONDUCTED AS APPROPRIATE. SINCE NO LOT NUMBER WAS PROVIDED, NO MANUFACTURING RECORD EVALUATION COULD BE PERFORMED. REASON FOR DEVICE EXPLANT AND/OR REOPERATION: RIGHT RUPTURE, AND CAPSULAR CONTRACTURE; BAKER GRADE III. D4: UDI: AS THE CATALOG/MODEL NUMBER WAS NOT PROVIDED, THE (01)GTIN IS NOT AVAILABLE. MENTOR IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH MENTOR HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, MENTOR, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, MENTOR, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).

Additional Manufacturer Narrative · 0

ON JANUARY 20, 2026, THE MENTOR FAILURE ANALYSIS LAB RECEIVED THE DEVICE FOR EVALUATION. PER DEVICE RECEIVED, FOLLOWING INFORMATION HAS BEEN UPDATED ON THIS FORM: FIELD D1 FOR BRAND NAME HAS BEEN UPDATED TO "MENTOR MEMORYGEL BREAST IMPLANT". FIELD D4 FOR CATALOG NUMBER HAS BEEN UPDATED TO "3503751BC". LOT NUMBER "6974321"HAS BEEN ADDED TO FIELD D4. FIELD D4 FOR UNIQUE IDENTIFIER( UDI) HAS BEEN UPDATED TO (B)(4). D4: UDI: AS THE SERIAL NUMBER FOR THE DEVICE INVOLVED IN THE EVENT WAS NOT PROVIDED, THE FULL UDI IS CURRENTLY NOT AVAILABLE. MENTOR IS AWARE THAT THE DATE OF IMPLANT IS PRIOR TO THE MANUFACTURING DATE OF REPORTED LOT NUMBER 6974321. MULTIPLE FOLLOW UPS WERE COMPLETED FOR CLARIFICATION. HOWEVER, NO RESPONSE WAS RECEIVED. IF ADDITIONAL INFORMATION BECOMES AVAILABLE, IT WILL BE UPDATED AND SUPPLEMENTAL REPORT WILL BE SUBMITTED. ON JANUARY 27, 2026, DEVICE EVALUATION WAS COMPLETED AS FOLLOWS: MENTOR CONDUCTED A VISUAL INSPECTION, MICROSCOPIC EXAMINATION, AND SHELL THICKNESS OF THE RETURNED DEVICE. VISUAL ANALYSIS OF THE RETURNED DEVICE REVEALED THAT THE BREAST IMPLANT WAS FOUND TO BE RUPTURED. MICROSCOPIC EXAMINATION WAS PERFORMED, AND THE CAUSE OF THE TEAR COULD NOT BE IDENTIFIED. THEREFORE A THICKNESS MEASUREMENT WAS CONDUCTED ON THE SHELL AT THE TEAR, AND THE THICKNESS WAS WITHIN MANUFACTURING SPECIFICATIONS. MENTOR ALSO PERFORMED A MANUFACTURING RECORD EVALUATION RELATED TO THE REPORTED COMPLAINT FOR THE FINISHED DEVICE LOT NUMBER. NO MANUFACTURING NON-CONFORMANCES WERE IDENTIFIED AS PART OF THIS EVALUATION. ALTHOUGH NO CONCLUSION COULD BE REACHED ON THE CAUSE OF THE REPORTED EVENT, THE INSTRUCTIONS FOR USE CONTAIN THE FOLLOWING CAUTION: RUPTURE CAN OCCUR AT ANY TIME AFTER IMPLANTATION BUT IS MORE LIKELY TO OCCUR THE LONGER THE IMPLANT IS IMPLANTED. THE FOLLOWING MAY CAUSE IMPLANT TO RUPTURE: STRESSING THE IMPLANT DURING IMPLANTATION AND WEAKENING IT; FOLDING OR WRINKLING OF THE IMPLANT SHELL; EXCESSIVE FORCE TO THE CHEST (E.G. DURING CLOSED CAPSULOTOMY); TRAUMA; COMPRESSION DURING MAMMOGRAPHIC IMAGING; AND SEVERE CAPSULAR CONTRACTURE. BREAST IMPLANTS MAY ALSO SIMPLY WEAR OUT OVER TIME. REGARDING THE REPORTED CAPSULAR CONTRACTURE CONDITION, THERE ARE NOT ENOUGH DETAILS TO DETERMINE THE FACTORS THAT MAY HAVE CAUSED THIS CONDITION. CAPSULAR CONTRACTURE IN THE PATIENT¿S BREAST MIGHT BE THE RESULT OF THE BODY¿S INDIVIDUAL PHYSIOLOGICAL RESPONSE TO THE IMPLANTATION OF A FOREIGN OBJECT IN SOFT TISSUE. FURTHER, CAPSULAR CONTRACTURE IS A KNOWN COMPLICATION ASSOCIATED WITH THESE DEVICES AND IS REFERENCED IN OUR CURRENT PRODUCT INSERT DATA SHEET. AS PART OF MENTOR'S QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED ACCORDING TO APPROVED SPECIFICATIONS. BASED ON THE RESULTS OF THIS INVESTIGATION, IT WAS DETERMINED THAT THE PRODUCT MET THE MANUFACTURING RELEASE CRITERIA: NO CORRECTIVE AND PREVENTIVE ACTION (CAPA) IS REQUIRED NOW. ADDITIONAL COMPLAINT INFORMATION MONITORING FOR POTENTIAL SAFETY SIGNALS IS CONDUCTED THROUGH COMPLAINT TRENDING AS PART OF POST MARKET SURVEILLANCE. THIS SUPPLEMENTAL MEDWATCH REPORT IS FOR THE PATIENT¿S RIGHT-SIDED DEVICE. THE MANUFACTURER'S REFERENCE NUMBER: (B)(4).

Description of Event or Problem · 0

IT WAS REPORTED THAT A 51-YEAR-OLD CAUCASIAN FEMALE PATIENT UNDERWENT PRIMARY BREAST AUGMENTATION WITH UNKNOWN SIZE UNKNOWN GEL IMPLANTS AND EXPERIENCED BREAST IMPLANT RUPTURE AND CAPSULAR CONTRACTURE; BAKER GRADE III ON HER RIGHT SIDE POSTOPERATIVELY. THE PATIENT UNDERWENT BILATERAL REPLACEMENT SURGERY WITH SERIAL NUMBERS (B)(6) ON THE LEFT SIDE, AND WITH (B)(6) ON THE RIGHT SIDE ON (B)(6) 2025.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
28995 MENTOR MEMORYGEL BREAST IMPLANT PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED FTR MENTOR TEXAS 6974321 81317000310

Patients

Seq Age Sex Outcome Treatment
1 52 YR Female Required Intervention