FDA Adverse Event Malfunction Summary report: N

UNICEL DXC 800 SYNCHRON SYSTEM

MDR report key: 2395337 · Received January 3, 2012

Report

Report Number
2050012-2011-08517
Event Type
Malfunction
Date Received
January 3, 2012
Date of Event
December 7, 2011
Report Date
December 7, 2011
Manufacturer
BECKMAN COULTER, INC.
Product Code
JJE
PMA / PMN Number
K042291
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

BEC IDENTIFIER FOR THIS COMPLAINT IS (B)(4).

Description of Event or Problem · 1

CUSTOMER REPORTED TO BECKMAN COULTER, INC. (BEC) THAT THE MODULAR CHEMISTRY (MC) SAMPLE 3 WAY VALVE OF THE UNICEL DXC 800 SYNCHRON SYSTEM WAS LEAKING. CUSTOMER REPORTED THAT ERRONEOUS RESULTS WERE NOT GENERATED. THERE WAS NO REPORT OF ANY ADVERSE EVENT OR INJURY REQUIRING MEDICAL INTERVENTION OR PATIENT TREATMENT. BEC FIELD SERVICE ENGINEER (FSE) REPLACED THE DRIPPING T-VALVE. THE REPLACEMENT OF THE T-VALVE RESOLVED THE ISSUE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 UNICEL DXC 800 SYNCHRON SYSTEM ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE JJE BECKMAN COULTER, INC. DXC 800 PRO NA

Patients

Seq Age Sex Outcome Treatment
1