FDA Adverse Event
Malfunction
Summary report: N
UNICEL DXC 800 SYNCHRON SYSTEM
MDR report key: 2395337
·
Received January 3, 2012
Report
- Report Number
- 2050012-2011-08517
- Event Type
- Malfunction
- Date Received
- January 3, 2012
- Date of Event
- December 7, 2011
- Report Date
- December 7, 2011
- Manufacturer
- BECKMAN COULTER, INC.
- Product Code
- JJE
- PMA / PMN Number
- K042291
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
BEC IDENTIFIER FOR THIS COMPLAINT IS (B)(4).
Description of Event or Problem · 1
CUSTOMER REPORTED TO BECKMAN COULTER, INC. (BEC) THAT THE MODULAR CHEMISTRY (MC) SAMPLE 3 WAY VALVE OF THE UNICEL DXC 800 SYNCHRON SYSTEM WAS LEAKING. CUSTOMER REPORTED THAT ERRONEOUS RESULTS WERE NOT GENERATED. THERE WAS NO REPORT OF ANY ADVERSE EVENT OR INJURY REQUIRING MEDICAL INTERVENTION OR PATIENT TREATMENT. BEC FIELD SERVICE ENGINEER (FSE) REPLACED THE DRIPPING T-VALVE. THE REPLACEMENT OF THE T-VALVE RESOLVED THE ISSUE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | UNICEL DXC 800 SYNCHRON SYSTEM | ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE | JJE | BECKMAN COULTER, INC. | DXC 800 PRO | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |