FDA Adverse Event Malfunction Summary report: N

BD POSIFLUSH

MDR report key: 23949268 · Received January 2, 2026

Report

Report Number
1911916-2025-00824
Event Type
Malfunction
Date Received
January 2, 2026
Date of Event
December 10, 2025
Report Date
January 9, 2026
Manufacturer
BD MEDICAL (BD WEST) MEDICAL SURGICAL
Product Code
NGT
UDI-DI
00382903065479
PMA / PMN Number
K003553
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
PHARMACIST
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

(B)(4). FOLLOW UP FOR DEVICE EVALUATION. A BLACK PARTICULATE RESEMBLING AN INSECT WAS REPORTEDLY OBSERVED FLOATING IN THE SYRINGE. TO SUPPORT THE INVESTIGATION, ONE SAMPLE WITHOUT FLOW WRAP PACKAGING WAS RECEIVED FOR EVALUATION BY THE QUALITY TEAM. THE SYRINGE CONTAINED 3.5ML OF SOLUTION, AND THE PLUNGERS RUBBER STOPPER WAS POSITIONED AT THE 8ML GRADUATION MARK. A VISUAL INSPECTION IDENTIFIED NO FOREIGN MATTER. THE SOLUTION WAS SUBSEQUENTLY TRANSFERRED TO A GLASS CONTAINER, WHERE NO FOREIGN MATTER WAS OBSERVED UPON RE INSPECTION. A DEVICE HISTORY RECORD REVIEW WAS COMPLETED FOR MATERIAL NUMBER 306547, LOT 5212358. THE REVIEW REVEALED NO DETECTED QUALITY ISSUES DURING THE PRODUCTION OF THIS LOT THAT COULD HAVE CONTRIBUTED TO THE REPORTED CONDITION. THERE WERE NO RELATED QUALITY NOTIFICATIONS, AND ALL PROCESSES AND FINAL INSPECTIONS COMPLIED WITH SPECIFICATION REQUIREMENTS. TO DATE, NO OTHER SIMILAR EVENTS HAVE BEEN REPORTED FOR THIS LOT. BASED ON THE INVESTIGATION, INCLUDING THE ANALYSIS OF THE RETURNED SAMPLE, THE CUSTOMERS REPORTED CONDITION COULD NOT BE CONFIRMED.

Additional Manufacturer Narrative · 0

H.3. IF A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS COMPLETED, A SUPPLEMENTAL REPORT WILL BE FILED. B3. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD.

Description of Event or Problem · 0

NO HARM HAS COME TO THE PATIENT.

Description of Event or Problem · 0

IT WAS REPORTED THAT BD SYR 10ML PUMP COMPATIBLE SALINE 10ML FIL CONTAINED FOREIGN MATTER. VERBATIM: PATIENT STATES THERE IS SOMETHING BLACK FLOATING IN THE SYRINGE THAT LOOKS LIKE A BUG.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
4047 BD POSIFLUSH SALINE, VASCULAR ACCESS FLUSH NGT BD MEDICAL (BD WEST) MEDICAL SURGICAL 5212358 00382903065479

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown Other