FDA Adverse Event Malfunction Summary report: N

UNICEL DXC 600 SYNCHRON SYSTEM

MDR report key: 2393709 · Received January 2, 2012

Report

Report Number
2050012-2011-08612
Event Type
Malfunction
Date Received
January 2, 2012
Date of Event
December 1, 2011
Report Date
December 1, 2011
Manufacturer
BECKMAN COULTER, INC.
Product Code
JJE
PMA / PMN Number
K042291
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NH, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT ISSUE WAS RESOLVED AFTER THE FSE REPLACED THE ISE PREAMP BOARD AND DRAIN VALVE ASSEMBLY. AN ADDITIONAL MEDICAL DEVICE REPORT WAS FILED BASED ON THE SAME COMPLAINT AS MEDWATCH #2050012-2011-08613, BECKMAN COULTER, INC. REPORT IDENTIFIER (B)(4). (NOTE: THE BECKMAN COULTER, INC. IDENTIFIER FOR THIS REPORT IS (B)(4).)

Description of Event or Problem · 1

CUSTOMER CALLED TO REPORT THAT THE UNICEL DXC 600 SYNCHRON SYSTEM GENERATED FALSELY LOW SODIUM (NA) RESULTS. THE RESULTS WERE REPORTED OUTSIDE THE LABORATORY; HOWEVER, WHEN THE ISSUE WAS NOTICED, THE SAMPLES WERE REPEATED AFTER THE SYSTEM WAS RECALIBRATED, AND THE REPORTS WERE AMENDED. THEREFORE, THERE WAS NO AFFECT TO PATIENT TREATMENT. THE DIFFERENCES IN RESULTS FOR ALL OF THE AMENDED SAMPLES WERE WITHIN THE PRECISION CLAIMS OF THE ASSAY. THE CUSTOMER TECHNICAL SPECIALIST (CTS) SCHEDULED A SERVICE VISIT. ON THE FOLLOWING DAY, THE FIELD SERVICE ENGINEER (FSE) INSPECTED THE INSTRUMENT AND REPLACED THE ISE (ION SELECTIVE ELECTRODE) PREAMP BOARD AND THE DRAIN VALVE ASSEMBLY, AND CONFIRMED THAT THE ISE HEALTH CHECK TESTING PASSED SPECIFICATIONS. FINALLY, THE FSE VERIFIED INSTRUMENT PERFORMANCE TO ENSURE THAT THE TROUBLESHOOTING EFFECTIVELY RESOLVED THE INSTRUMENT ISSUE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 UNICEL DXC 600 SYNCHRON SYSTEM ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE JJE BECKMAN COULTER, INC. DXC 600 PRO

Patients

Seq Age Sex Outcome Treatment
1