FDA Adverse Event Malfunction Summary report: N

BECKMAN COULTER LAB AUTOMATION SYSTEM

MDR report key: 2393612 · Received January 2, 2012

Report

Report Number
2050012-2011-08605
Event Type
Malfunction
Date Received
January 2, 2012
Date of Event
December 1, 2011
Report Date
December 1, 2011
Manufacturer
BECKMAN COULTER BIOMEDICAL GMBH
Product Code
JQP
PMA / PMN Number
EXEMPT
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE CUSTOMER INDICATED THAT THE COVER ITSELF (AND THE GAS SPRINGS) HAVE AN ADDITIONAL LOCKING BRACKET WHERE ONCE THE COVER IS OPENED, WHEN PROPERLY ACTIVATED, PREVENTS A CRASHING DOWN OF THE COVER IN THE EVENT OF INADEQUATE STRENGTH OF THE SPRINGS. THE USER IS CAUTIONED IN THE USER GUIDE TO MAKE SURE THE LOCKS ARE IN PLACE WHEN OPENING THE COVER. HOWEVER, THE COVERS SHOULD NOT BE OPENED TO REMOVE OR INSERT RACKS INTO THE SYSTEM. THE OPERATOR VIOLATED THE INTENDED USE OF THE SYSTEM. THE OPERATOR DISREGARDED THE SAFETY INSTRUCTIONS AND WARNINGS GIVEN IN THE BEC IFU (INSTRUCTIONS FOR USE) MANUAL AS WELL AS THE TWO SAFETY LABELS ON THE COVER OF THE AUTOMATE SYSTEM AND DID NOT OPERATE THE SYSTEM AS SPECIFIED: IFU TELLS NOT TO REACH INTO THE DEVICE FOR LOADING/UNLOADING TRAYS. LOADING/UNLOADING TRAYS IS ONLY TO DO WITH CLOSED COVERS ON FULLY EXTENDED DRAWERS, COVERS ARE ONLY NEEDED TO BE OPENED IN CASE OF FAILURE WHICH NEEDS USER INTERACTION AND FOR CLEANING PURPOSE, THE NORMAL STATE OF THE COVERS IS CLOSED, ADDITIONALLY THE STICKER ON THE COVER ITSELF TELLS THE USER TO MAKE SURE THE COVER STAYS IN UP POSITION BEFORE REACHING INTO THE DEVICE. A BEC FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED ON (B)(6) 2011 AND PERFORMED TROUBLESHOOTING ON THE SYSTEM. THE FSE REPLACED BOTH GAS SPRINGS. PARTS WERE NOT AVAILABLE FOR ANY FURTHER INVESTIGATION. THE FSE ALSO INFORMED THE CUSTOMER TO NOT VIOLATE THE IFU AND THE SAFETY STICKERS ON THE COVER.

Description of Event or Problem · 1

A CUSTOMER CONTACTED BECKMAN COULTER INC. (BEC) STATING THAT DURING THE LOADING/UNLOADING PROCESS OF THE AUTOMATE 2500 HS S3I (SORTER AND ALIQUOTER) SYSTEM WITH TUBE RACKS THE COVER UNEXPECTEDLY CLOSED AND STRUCK THE OPERATOR ON THE HEAD. THE OPERATOR WAS HIT ON THE FOREHEAD AND ENDURED A HEAD CONTUSION (NO CUT, NO BRUISE, SLIGHT BUMP ON THE HEAD). NO MEDICAL INTERVENTION WAS NECESSARY. THE CUSTOMER INDICATED THAT THE ANALYZER WAS FITTED WITH OLD GAS SPRINGS, WHICH MAY BECOME WORN AND DETERIORATE OVER TIME. IT IS UNKNOWN WHETHER THE GAS SPRINGS ON THE COVER WERE THE ORIGINAL ONES OR HAD ALREADY BEEN REPLACED IN THE SYSTEMS HISTORY. NO PATIENT SAMPLES WERE INVOLVED IN THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 BECKMAN COULTER LAB AUTOMATION SYSTEM CALCULATOR/DATA PROCESSING MODULE, FOR CLINICAL USE JQP BECKMAN COULTER BIOMEDICAL GMBH AUTOMATE 2500 N/A

Patients

Seq Age Sex Outcome Treatment
1