FDA Adverse Event Malfunction Summary report: N

RESTORE ULTRA

MDR report key: 2392472 · Received December 29, 2011

Report

Report Number
3004209178-2011-10115
Event Type
Malfunction
Date Received
December 29, 2011
Report Date
December 8, 2011
Manufacturer
MDT PUERTO RICO OPERATIONS CO
Product Code
LGW
PMA / PMN Number
P840001
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
ND, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

PROGRAMMER: MODEL 37743, SERIAL# (B)(4). RECHARGER: MODEL 37752, SERIAL# (B)(4). LEAD: MODEL 3777-60, SERIAL# (B)(4), IMPLANTED: (B)(6) 2009, EXPLANTED: NA. LEAD: MODEL 3777-60, SERIAL#(B)(4), IMPLANTED: (B)(6) 2009, EXPLANTED: NA. ANCHOR: MODEL 3550-39, LOT# N213355.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT EXPERIENCED COUPLING ISSUES WITH THEIR IMPLANTABLE NEUROSTIMULATOR. AN ANTENNA LOCATOR WAS USED AND RESULTED IN A POWER-ON-RESET CONDITION. A NORMAL CHARGING SCREEN WAS REACHED AND THE PATIENT RECEIVED PERFECT COUPLING. ADDITIONAL INFORMATION HAD BEEN REQUESTED, BUT WAS NOT AVAILABLE AS OF THE DATE OF THIS REPORT.

Description of Event or Problem · 1

FOLLOW UP INFORMATION RECEIVED INDICATED THAT THE PATIENT'S PROGRAMMER WAS WORKING AND HAD NO ISSUES WITH ITS USE. PREVIOUSLY, SHE HAD LOST THE PROGRAMMER AND HAD TO BE SENT A NEW ONE. SHE STRUGGLES WITH THE USE OF THE EXTERNAL RECHARGER BUT WAS PERFORMING THE TASK CORRECTLY. DUE TO THE PATIENT'S SMALL SIZE SHE "STRUGGLED" TO MAINTAIN A GOOD CONNECTION DURING RECHARGING AND CONTINUALLY HAD TO REPOSITION HERSELF TO MAINTAIN THE HIGHEST LEVEL OF COUPLING. IT WAS ALSO NOTED THAT THE RECHARGER DID NOT LIE FLAT OVER HER IMPLANTABLE NEUROSTIMULATOR (INS) DUE TO HER SIZE AND INS PLACEMENT LOCATION. IF ADDITIONAL INFORMATION IS RECEIVED, A FOLLOW UP REPORT WILL BE SENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 RESTORE ULTRA STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF LGW MDT PUERTO RICO OPERATIONS CO 37712

Patients

Seq Age Sex Outcome Treatment