FDA Adverse Event Malfunction Summary report: N

PIPELINE VANTAGE WITH SHIELD TECHNOLOGY

MDR report key: 23924696 · Received December 30, 2025

Report

Report Number
2029214-2025-02699
Event Type
Malfunction
Date Received
December 30, 2025
Date of Event
December 22, 2025
Report Date
January 13, 2026
Manufacturer
MICRO THERAPEUTICS, INC. DBA EV3
Product Code
OUT
UDI-DI
00847536035195
PMA / PMN Number
P100018
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IT
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

B5 UPDATED WITH ADDITIONAL INFORMATION RECEIVED. H6. DEVICE CODE UPDATED BASED ON AVAILABLE INFORMATION. THIS DEVICE IS NO LONGER REPORTABLE AND IS NOT CONSIDERED A COMPLAINT AS THERE WAS NO DEVICE MALFUNCTION/DEFICIENCY AND NO PATIENT COMPLICATION OR INJURY ASSOCIATED WITH THIS EVENT. NO ADDITIONAL SUPPLEMENTAL MDRS ARE REQUIRED UNLESS ADDITIONAL INFORMATION RECEIVED MAKES THIS DEVICES REPORTABLE. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Additional Manufacturer Narrative · 0

CONTINUATION OF D10: PRODUCT ID PED3-021-350-20 (D050452); PRODUCT TYPE: ; IMPLANT DATE N/A; EXPLANT DATE N/A PRODUCT ID FG13150-0615-2S (229938289); PRODUCT TYPE: ; IMPLANT DATE N/A; EXPLANT DATE N/A PRODUCT ID FG19120-1030-1S (231481250); PRODUCT TYPE: ; IMPLANT DATE N/A; EXPLANT DATE N/A MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Description of Event or Problem · 0

ADDITIONAL INFORMATION RECEIVED REPORTED THE RESHEATHING ISSUE OCCURRED ONLY WITH THE FIRST PIPELINE (PED3-021-350-20. LOT:D050452) AND THE PHENOM 21 MICRO CATHETER (FG13150-0615-2S, LOT: 229938289). RESISTANCE OCCURRED WITH THE PIPELINE IN THE DISTAL END OF THE PHENOM 21 MICROCATHETER. IT WAS NOTED THAT CONTINUOUS FLUSH WAS ADMINISTERED AND MAINTAINED DURING THE PROCEDURE AS INDICATED IN THE IFU. THERE WAS NO DAMAGE OBSERVED TO THE PIPELINE PUSHWIRE, NOR TO THE PHENOM21 OR PHENOM PLUS CATHETERS. THE SECOND PIPELINE WAS SUCCESSFULLY IMPLANTED TO COMPLETE THE PROCEDURE. THERE WAS NO ISSUE REPORTED INVOLVING THE PHENOM PLUS CATHETER (FG19120-1030-1S, LOT: 231481250) OR THE SECOND/IMPLANTED PIPELINE (PED3-021-350-16, LOT: D050415).

Description of Event or Problem · 0

IT WAS REPORTED THAT DURING THE DEPLOYMENT OF THE PIPELINE VANTAGE WITH SHIELD TECHNOLOGY FLOW DIVERTER DEVICE FOR THE TREATMENT OF AN UNRUPTURED, FUSIFORM CAROTID TIP ANEURYSM (MAXIMUM DIAMETER 5 MM, NECK DIAMETER 4 MM), IT WAS NOT POSSIBLE TO RESHEATH THE DEVICE DURING THE PROCEDURE, DESPITE NO VISUAL EVIDENCE OF TWISTING. MULTIPLE DEVICES WERE INVOLVED. ATTEMPTS TO RESHEATH THE DEVICE OUTSIDE THE SYSTEM WERE ALSO UNSUCCESSFUL, AND UPON OPENING, INCREASED RADIOPACITY WAS OBSERVED AT THE ZONE INVOLVED IN THE UNSUCCESSFUL RESHEATHING PROCESS. THE PROCEDURE WAS PERFORMED USING TRIAXIAL ACCESS VIA THE FEMORAL ARTERY (DIAMETER 8 MM) WITH NORMAL VESSEL TORTUOSITY, AND DUAL ANTIPLATELET THERAPY (ASA AND PLAVIX) WAS ADMINISTERED. GUIDE CATHETER (INFINITY PLUS) AND GUIDEWIRE (SYNCRO) WERE USED, AND ALL DEVICES AND ACCESSORIES WERE PREPARED AND FLUSHED AS INDICATED IN THE INSTRUCTIONS FOR USE. THE ANGIOGRAPHIC RESULT POST PROCEDURE WAS REPORTED AS GOOD. NO PATIENT SYMPTOMS OR COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT. THE LANDING ZONE WAS 3.3MM DISTAL AND 3.5MM PROXIMAL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2259422 PIPELINE VANTAGE WITH SHIELD TECHNOLOGY INTRACRANIAL ANEURYSM FLOW DIVERTER OUT MICRO THERAPEUTICS, INC. DBA EV3 PED3-021-350-16 D050415 00847536035195

Patients

Seq Age Sex Outcome Treatment
1 54 YR Male