COD.712 TOP CIRC.RACC.CARDIO
Report
- Report Number
- 1718850-2011-00252
- Event Type
- Other
- Date Received
- December 15, 2011
- Date of Event
- November 10, 2011
- Report Date
- November 14, 2011
- Manufacturer
- SORIN GROUP ITALIA
- Product Code
- DTN
- PMA / PMN Number
- K992599
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- OTHER
Narratives
SORIN GROUP (B)(4) MANUFACTURES THIS CARDIOTOMY RESERVOIR. ALTHOUGH CAT # 0405 IS NOT MARKETED IN THE U.S, IT IS EQUIVALENT TO US CAT # 04116 WHICH WAS CLEARED UNDER 510(K)# K992599. THE INCIDENT OCCURRED IN (B)(6). THIS MEDWATCH IS BEING FILED ON BEHALF OF SORIN GROUP (B)(4). SORIN GROUP (B)(4) RECEIVED A REPORT THAT THE FILTER BECAME DETACHED FROM THE AUTO-TRANSFUSION RESERVOIR. THE DETACHMENT OF THE FILTER ALLOWED DEBRIS TO PASS INTO THE AUTO-TRANSFUSION MACHINE CAUSING A BLOCKAGE AND PRESSURE INCREASE. THE INCREASED PRESSURE CAUSED THE WASH BOWL TO FRACTURE, RESULTING IN AN INABILITY TO RE-INFUSE THE SALVAGED BLOOD. THERE AS NO REPORT OF PT INJURY. IT WAS LATER REPORTED THAT THE PT HAD RARE ANTIBODIES SO BANK BLOOD WAS NOT AVAILABLE AND SPECIAL PROCEDURES HAD TO IMPLEMENTED TO PROVIDE THE NECESSARY BLOOD PRODUCTS. THE DEVICE WAS DISCARDED BY THE USER AND IS, THEREFORE, NOT AVAILABLE FOR EVAL. WITHOUT THE RESERVOIR FOR EVAL, THE ROOT CAUSE FOR THE ISSUE CANNOT BE DETERMINED. THERE WAS NO REPORT OF PT INJURY AS A RESULT OF THIS EVENT. THE ELECTA AUTOTRANSFUSION MACHINE INSTRUCTIONS FOR USE STATE, "AN ADEQUATE QUANTITY OF SALVAGED, VIABLE, WASHED, PACKED RED BLOOD CELLS CANNOT ALWAYS BE PREDICTED OR DEPENDED UPON IN A BLOOD SALVAGE PROCEDURE. PHYSICIAN SHOULD BE PREPARED TO INSTITUTE APPROPRIATE ADD'L THERAPY IF NECESSARY."
SORIN GROUP (B)(4) RECEIVED A REPORT THAT THE FILTER BECAME DETACHED FROM THE AUTO-TRANSFUSION RESERVOIR. THE DETACHMENT OF THE FILTER ALLOWED DEBRIS TO PASS INTO THE AUTO-TRANSFUSION MACHINE CAUSING A BLOCKAGE AND PRESSURE INCREASE. THE INCREASED PRESSURE CAUSED THE WASH BOWL TO FRACTURE, RESULTING IN AN INABILITY TO RE-INFUSE THE SALVAGED BLOOD. THERE WAS NO REPORT OF PT INJURY..
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COD.712 TOP CIRC.RACC.CARDIO | AUTOTRANSFUSION UNIT, BLOOD COLLECTION | DTN | SORIN GROUP ITALIA | NA | 1106220055 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |