FDA Adverse Event Malfunction Summary report: N

COOLTOUCH MODEL 500R FIBER OPTIC

MDR report key: 2392264 · Received November 29, 2007

Report

Report Number
2951571-2007-00001
Event Type
Malfunction
Date Received
November 29, 2007
Date of Event
November 15, 2007
Report Date
November 29, 2001
Manufacturer
NEW STAR LASERS, INC., DBA COOLTOUCH, INC.
Product Code
GEX
Product Problem
Yes
Report Source
Manufacturer report

Narratives

Additional Manufacturer Narrative · 1

COMPANY REP CONFIRMED "GREEN STICK" BREAK (FIBER OPTIC BROKE DUE TO EXCESS PRESSURE EXERTED ON IT). FIBER WAS REPLACED.

Description of Event or Problem · 1

THE PHYSICIAN WAS PERFORMING CONVENTIONAL LIPOSUCTION ON THE PT'S ABDOMEN IN CONJUNCTION WITH A 1320 NM LASER. DURING THE PROCESS OF APPLYING THERMAL HEATING IN A STANDARD MANNER AND AT THE POINT WHERE THE FIBER OPTIC EXITS THE HANDPIECE, THE FIBER HAD PUSHED AGAINST THE DOCTOR AND THE FIBER BROKE UNNOTICED. WITH THE LASER STILL FIRING, THE BROKEN FIBER PIECE FELL TO THE FLOOR, BRUSHING THE PAPER DRAPES AND IGNITING THEM. THE PHYSICIAN AND ATTENDANT ATTEMPTED TO EXTINGUISH THE FLAMES BY SMOTHERING THEM BUT WERE UNSUCCESSFUL. STAFF GRABBED A FIRE EXTINGUISHER AND THE FIRE WAS EXTINGUISHED. THE PT WAS UNHARMED AND REMOVED TO ANOTHER ROOM, WHERE A NEW FIBER WAS USED AND THE PROCEDURE WAS COMPLETED WITH NO ISSUES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COOLTOUCH MODEL 500R FIBER OPTIC COOLTOUCH FIBER OPTIC GEX NEW STAR LASERS, INC., DBA COOLTOUCH, INC.

Patients

Seq Age Sex Outcome Treatment
1 45 YR