ARCHITECT HBSAG
Report
- Report Number
- 3008344661-2025-00191
- Event Type
- Malfunction
- Date Received
- December 30, 2025
- Date of Event
- December 17, 2025
- Report Date
- February 4, 2026
- Manufacturer
- ABBOTT IRELAND DIAGNOSTICS DIVISION
- Product Code
- LOM
- UDI-DI
- 00380740167707
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
DATA AND INFORMATION PROVIDED WERE REVIEWED AND SUPPORT THE COMPLAINT ISSUE. REVIEW OF TRACKING AND TRENDING FOR THE ARCHITECT HBSAG ASSAY DID NOT IDENTIFY ANY TRENDS RELATED TO THE COMPLAINT ISSUE. A DEVICE HISTORY RECORD REVIEW DID NOT IDENTIFY ANY NONCONFORMANCES, POTENTIAL NONCONFORMANCES OR DEVIATIONS ASSOCIATED WITH THE ARCHITECT HBSAG ASSAY AND COMPLAINT ISSUE. LABELING WAS REVIEWED WHICH ADEQUATELY ADDRESSES THE CURRENT ISSUE. REVIEW OF THE PUBLICATIONS, H. W. REESINK ET AL. ¿OCCULT HEPATITIS B INFECTION IN BLOOD DONORS¿, VOX SANGUINIS (2008) 94, 153-166 AND MICHAEL TORBENSON AND DAVID L THOMAS ¿OCCULT HEPATITIS B¿, LANCET INFECT DIS 2002; 2: 479-86, THIS COULD POTENTIALLY BE A CASE OF OCCULT HBV INFECTION WHICH IS A KNOWN PHENOMENON AND IS DIAGNOSED WHEN A HBV DNA TEST IS POSITIVE BUT HBSAG IS UNDETECTED. OCCULT INFECTION MAY REPRESENT ACUTE INFECTION IN THE WINDOW PERIOD, HBV TAIL END OF CHRONIC HBV INFECTION, PERSISTENCE OF REPLICATION AT LOW LEVEL AFTER RECOVERY IN THE PRESENCE OF ANTI-HBS OR OCCURRENCE OF AN ESCAPE MUTANT IN VACCINATED OR UNVACCINATED INDIVIDUALS NOT DETECTED BY CURRENT HBSAG ASSAYS. BASED ON THE INVESTIGATION, NO SYSTEMIC ISSUE OR DEFICIENCY OF THE ARCHITECT HBSAG REAGENT WAS IDENTIFIED.
THIS REPORT IS BEING FILED ON AN INTERNATIONAL PRODUCT, LIST NUMBER 6C36 THAT HAS A SIMILAR PRODUCT DISTRIBUTED IN THE US, LIST NUMBER 4P53, AND 510K/PMA/BLA OF P110029. SECTION A PATIENT INFORMATION: NO ADDITIONAL INFORMATION PROVIDED. AN EVALUATION IS IN PROCESS. A FOLLOWUP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE.
THE CUSTOMER OBSERVED FALSE NEGATIVE ARCHITECT HBSAG RESULT. A PATIENT WITH CHRONIC HEPATITIS B UNDERGOING ANTIVIRAL THERAPY HAD A NEGATIVE HBSAG RESULT BUT A WEAKLY POSITIVE HBV-DNA RESULT. THE MEDICAL HISTORY IS HBSAG OF 0.2 IU/ML, REACTIVE. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
THE CUSTOMER OBSERVED FALSE NEGATIVE ARCHITECT HBSAG RESULT. A PATIENT WITH CHRONIC HEPATITIS B UNDERGOING ANTIVIRAL THERAPY HAD A NEGATIVE HBSAG RESULT BUT A WEAKLY POSITIVE HBV-DNA RESULT. THE MEDICAL HISTORY IS HBSAG OF 0.2 IU/ML, REACTIVE. NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1835756 | ARCHITECT HBSAG | TEST, HEPATITIS B (B CORE, BE ANTIGEN, BE ANTIBODY, B CORE IGM) | LOM | ABBOTT IRELAND DIAGNOSTICS DIVISION | 00380740167707 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 41 YR | Male | ARC I2000SR INST, 03M74-02, (B)(6) |