FDA Adverse Event Malfunction Summary report: N

CD HORIZON® SPINAL SYSTEM

MDR report key: 23919261 · Received December 29, 2025

Report

Report Number
3003120897-2025-00819
Event Type
Malfunction
Date Received
December 29, 2025
Date of Event
December 8, 2025
Report Date
February 4, 2026
Manufacturer
MEDTRONIC SOFAMOR DANEK USA, INC.
Product Code
HXX
UDI-DI
00643169043640
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WA, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

PRODUCT ANALYSIS # (B)(4), PART # 5584111, LOT # K24G1499 VISUAL INSPECTION CONFIRMED THE ENTIRE TORX TIP OF INSTRUMENT HAS BEEN SHEARED OFF AND THE ATTACHMENT PIN TO THE INTERNAL BUSHING HAS BROKEN. OPTICAL EXAMINATION OF THE FRACTURE SURFACE REVEALED A FAIRLY FLAT FRACTURE SURFACE AND CIRCULAR MATERIAL FLOW. THE DAMAGE TO THE DRIVER IS CONSISTENT OVERLOAD. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Additional Manufacturer Narrative · 0

MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Description of Event or Problem · 0

INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAVING TRANSFORAMINAL LUMBAR INTERBODY FUSION SPINAL THERAPY TO L5-S1. IT WAS REPORTED THAT THE TIP OF THE SCREWDRIVER BROKE WHILE THE SURGEON WAS PLACING THE L5 SCREW ON THE PATIENT'S LEFT SIDE. THE BROKEN TIP WAS DISCARDED IN THE TRASH AT THE END OF THE CASE. NO FRAGMENTS REMAINED IN THE PATIENT. THERE WERE NO PATIENT SYMPTOMS OR COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
363015 CD HORIZON® SPINAL SYSTEM SCREWDRIVER HXX MEDTRONIC SOFAMOR DANEK USA, INC. 5584111 K24G1499 00643169043640

Patients

Seq Age Sex Outcome Treatment
1 66 YR Female