Additional Manufacturer Narrative · 0

ONE POWER SUPPLY AND ONE POWER CORD MODEL WAS RECEIVED FOR EVALUATION. VISUAL INSPECTION VERIFIED THE POWER SUPPLY WAS FRAYED AND WIRES WERE EXPOSED. THE DEVICE HISTORY RECORD WAS REVIEWED TO ENSURE THAT EACH MANUFACTURING AND INSPECTION OPERATION WAS PERFORMED. THE REVIEW DETERMINED THAT NO NON-CONFORMANCES WERE IDENTIFIED THAT ARE RELATED TO THE REPORTED EVENT.

Additional Manufacturer Narrative · 0

THE INVESTIGATION CODES ALONG WITH INVESTIGATION RESULTS WILL BE PROVIDED IN A SUBSEQUENT SUBMISSION.

Description of Event or Problem · 0

THE PATIENT REPORTED SPARKING ON THE SUPPLY. THE PATIENT ELECTRONIC UNIT WAS REPLACED AND THERE WERE NO ADVERSE CONSEQUENCES REPORTED BY THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1893106 CARDIOMEMS POWER SUPPLY SYSTEM, HEMODYNAMIC, IMPLANTABLE MOM ABBOTT MEDICAL. CM1110

Patients

Seq Age Sex Outcome Treatment
1 75 YR Female