FDA Adverse Event Malfunction Summary report: N

ELECSYS ACTH

MDR report key: 23905893 · Received December 29, 2025

Report

Report Number
1823260-2025-05351
Event Type
Malfunction
Date Received
December 29, 2025
Date of Event
December 9, 2025
Report Date
December 29, 2025
Manufacturer
ROCHE DIAGNOSTICS
Product Code
CKG
UDI-DI
04015630939893
PMA / PMN Number
K060585
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE E801 ANALYZER SERIAL NUMBER WAS (B)(6). SAMPLE MATERIAL FROM THE PATIENT WAS RECEIVED FOR INVESTIGATION. THE CUSTOMER'S HIGH ACTH RESULTS WERE REPRODUCED AT THE INVESTIGATION SITE. SAMPLE MATERIAL FROM THE PATIENT WAS SENT TO AN EXTERNAL LABORATORY AND TESTED BY THE SIEMENS METHOD, WHERE THE ACTH WAS ALSO EXTREMELY HIGH (80,163 PG/ML). INTERFERENCE TESTING USING SIZE EXCLUSION CHROMATOGRAPHY (SEC) WAS PERFORMED. NO INTERFERENCE WAS FOUND. PEPTIDE MAPPING ANALYSIS OF THE SAMPLE FRACTIONS SUGGESTS THE HIGH ELECSYS ACTH RESULTS ARE CONSISTENT WITH AN ERRONEOUS DETECTION OF A DIFFERENT PEPTIDE FROM THE HIGHER-LEVEL PROOPIOMELANOCORTIN (POMC) PROTEIN.

Description of Event or Problem · 0

THERE WAS AN ALLEGATION OF QUESTIONABLE HIGH RESULTS FOR 1 PATIENT TESTED FOR ELECSYS ACTH (ACTH) ON A COBAS E 801 ANALYTICAL UNIT. THE ACTH RESULTS ARE ALLEGEDLY HIGH WITH NO KNOWN CAUSE, AS THE PATIENT IS ASYMPTOMATIC (PREVIOUSLY CUSHING'S DISEASE). THE INITIAL RESULT WAS 118,000 PG/ML. THE SAMPLE WAS REPEATED WITH VARIOUS DILUTIONS: X5: > 2000 PG/ML WITH A DATA FLAG, X10: >2000 PG/ML WITH A DATA FLAG, X20: >2000 PG/ML WITH A DATA FLAG, X40: >2000 PG/ML WITH A DATA FLAG, X75: 118,350 PG/ML, X100: 126,400 PG/ML. ON (B)(6) 2025, A NEW SAMPLE WAS OBTAINED, AND THE RESULT WAS APPROXIMATELY 25,000 PG/ML.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2847947 ELECSYS ACTH ADRENOCORTICOTROPIC HORMONE (ACTH) TEST SYSTEM CKG ROCHE DIAGNOSTICS ASKU 04015630939893

Patients

Seq Age Sex Outcome Treatment
1 65 YR Female DETA| ESCITALOPRAM| ESTRADIOL| FISH OIL| FLUDROCORTIZONE| HYDROCORTISONE| PREDNISONE| PROBIOTICS| PSYLLIUM| SOLU-CORTEF| SOMATROPIN INJECTION| VITAMIN C