FDA Adverse Event Malfunction Summary report: N

CARDIOMEMS POWER SUPPLY

MDR report key: 23888341 · Received December 24, 2025

Report

Report Number
3004936110-2025-06637
Event Type
Malfunction
Date Received
December 24, 2025
Date of Event
December 19, 2025
Report Date
March 31, 2026
Manufacturer
ABBOTT MEDICAL
Product Code
MOM
PMA / PMN Number
P100045
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

ONE POWER SUPPLY AND ONE POWER CORD MODEL WAS RECEIVED FOR EVALUATION. VISUAL INSPECTION VERIFIED THE POWER SUPPLY WAS FRAYED AND WIRES WERE EXPOSED. THE DEVICE HISTORY RECORD WAS REVIEWED TO ENSURE THAT EACH MANUFACTURING AND INSPECTION OPERATION WAS PERFORMED. THE REVIEW DETERMINED THAT NO NON-CONFORMANCES WERE IDENTIFIED THAT ARE RELATED TO THE REPORTED EVENT.

Additional Manufacturer Narrative · 0

THE INVESTIGATION CODES ALONG WITH INVESTIGATION RESULTS WILL BE PROVIDED IN A SUBSEQUENT SUBMISSION.

Description of Event or Problem · 0

THE PATIENT REPORTED EXPOSED AND FRAYED WIRES OF THE POWER CORD. THE POWER CORD AND POWER SUPPLY WERE REPLACED AND THERE WERE NO ADVERSE CONSEQUENCES REPORTED BY THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2476972 CARDIOMEMS POWER SUPPLY SYSTEM, HEMODYNAMIC, IMPLANTABLE MOM ABBOTT MEDICAL CM1110

Patients

Seq Age Sex Outcome Treatment
1 71 YR Female