STYLE 68 SALINE FILLED BREAST IMPLANT
Report
- Report Number
- 9617229-2025-22519
- Event Type
- Injury
- Date Received
- December 23, 2025
- Date of Event
- December 5, 2025
- Report Date
- April 22, 2026
- Manufacturer
- ALLERGAN (COSTA RICA)
- Product Code
- FWM
- PMA / PMN Number
- P990074
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
A REVIEW OF THE DEVICE HISTORY RECORD HAS BEEN COMPLETED. NO DEVIATIONS OR NON-CONFORMANCES NOTED. FURTHER INFORMATION FROM THE REPORTER REGARDING EVENT, PRODUCT, OR PATIENT DETAILS HAS BEEN REQUESTED. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME. REASON FOR REOPERATION: DEFLATION.
ADDITIONAL, CHANGED, AND/OR CORRECTED DATA: A2
DEVICE EVALUATION: THE DEVICE RELATED TO THE REPORTED EVENTS DEFLATION WAS RECEIVED ON MARCH 27, 2026, WITH LOT NUMBER 2925530. PER THE INVESTIGATION PROCEDURE, THE DEVICE IS ANALYZED THROUGH VISUAL INSPECTION MICROSCOPIC INSPECTION IF OPENINGS ARE OBSERVED AND A WEIGHT VERIFICATION (IF APPLICABLE). PER THE ANALYSIS PERFORMED, THE ASSESSMENTS OF THE COMPLAINTS AND ANY POTENTIAL MANUFACTURING ISSUE ARE DISPLAYED ALONG WITH ANY FURTHER ACTIONS REQUIRED: ¿ DEFLATION: OBSERVED AN OPENING THROUGH MICROSCOPIC INSPECTION ASSESSED AS UNIDENTIFIED (TEAR) OPENING. NO FURTHER ACTIONS ARE REQUIRED AS IT IS NOT RELATED TO THE MANUFACTURING PROCESS. NONE OF THE OTHER OBSERVATIONS PERFORMED DURING THE DEVICE ANALYSIS (CREASES, WEAR ABRASION) ARE FOUND TO BE POTENTIALLY RELATED TO THE MANUFACTURING PROCESS, AND, THEREFORE, NO FURTHER ACTIONS ARE REQUIRED FOR THESE OBSERVATIONS.
ADDITIONAL, CHANGED, AND/OR CORRECTED DATA: B5, D6B, H6
HEALTHCARE PROFESSIONAL REPORTED "DEFLATION". THIS RECORD IS FOR THE RIGHT SIDE. DEVICE REMAINS IMPLANTED.
HEALTHCARE PROFESSIONAL REPORTED "DEFLATION". THIS RECORD IS FOR THE RIGHT SIDE. DEVICE WAS EXPLANTED.
HEALTHCARE PROFESSIONAL REPORTED "DEFLATION". THIS RECORD IS FOR THE RIGHT SIDE. DEVICE WAS EXPLANTED.
DEVICE HAS BEEN EXPLANTED AND REPLACED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 80580 | STYLE 68 SALINE FILLED BREAST IMPLANT | PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE | FWM | ALLERGAN (COSTA RICA) | 2925530 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 65 YR | Female | Required Intervention |