UNICEL DXC 600I SYNCHRON ACCESS CLINICAL SYSTEM
Report
- Report Number
- 2050012-2011-08540
- Event Type
- Malfunction
- Date Received
- December 23, 2011
- Date of Event
- November 27, 2011
- Report Date
- November 27, 2011
- Manufacturer
- BECKMAN COULTER, INC.
- Product Code
- JJE
- PMA / PMN Number
- K060256
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NM, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
THE BECKMAN COULTER, INC. IDENTIFIER FOR THIS REPORT IS (B)(4).
CUSTOMER CALLED TO REPORT THAT FLUID WAS COLLECTING UNDERNEATH THE CARTRIDGE CHEMISTRY (CC) SAMPLE PROBE OF THE UNICEL DXC 600I SYNCHRON ACCESS CLINICAL SYSTEM WHEN IN HOME POSITION. THE CUSTOMER TECHNICAL SPECIALIST (CTS) IDENTIFIED THE LEAKING FLUID AS EITHER DI H20 (REAGENT CONTAINER RINSE) OR WASH SOLUTION, THEN SCHEDULED A SERVICE VISIT. ON THE FOLLOWING DAY, THE FIELD SERVICE ENGINEER (FSE) INSPECTED THE INSTRUMENT AND FOUND THAT THE CC SAMPLE PROBE WAS DRIPPING FLUID. THE FSE TROUBLESHOT THE INSTRUMENT BY REMOVING THE WASH LINE FROM THE COLLAR WASH, AND THEN BY PRIMING THE INSTRUMENT. THE FSE THEN SWAPPED THE T-VALVES FROM THE MODULAR CHEMISTRY (MC) SYRINGE TO THE CC SYRINGE. THE DRIPPING CEASED ON THE CC SAMPLE PROBE; HOWEVER, AFTER PRIMING, THE MC SAMPLE PROBE WAS DRIPPING FLUID. THE FSE DETERMINED THAT THE ROOT CAUSE OF THE ISSUE WAS DUE TO A T-VALVE ISSUE. THE FSE ORDERED THE REQUIRED REPLACEMENT PARTS AND RETURNED ON THE FOLLOWING DAY TO REPLACE THE T-VALVE, AFTER WHICH NO FURTHER LEAKING WAS OBSERVED. AFTER TESTING QUALITY CONTROLS, THE FSE VERIFIED INSTRUMENT PERFORMANCE TO ENSURE THAT THE SERVICES PERFORMED EFFECTIVELY RESOLVED THE INSTRUMENT ISSUE. CUSTOMER STATED THAT NO ERRONEOUS PATIENT RESULTS WERE GENERATED AND THAT NO ONE WAS HARMED BY OR EXPOSED TO THE LEAK.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | UNICEL DXC 600I SYNCHRON ACCESS CLINICAL SYSTEM | ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE | JJE | BECKMAN COULTER, INC. | DXC 600I |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |