FDA Adverse Event Malfunction Summary report: N

LIPOASPIRATE WASH SYSTEM

MDR report key: 23877800 · Received December 23, 2025

Report

Report Number
1651189-2025-09975
Event Type
Malfunction
Date Received
December 23, 2025
Date of Event
December 2, 2025
Report Date
December 2, 2025
Manufacturer
TIGER AESTHETICS MEDICAL
Product Code
QKL
PMA / PMN Number
BK190433
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

TIGER AESTHETICS MEDICAL COMPLAINT # (B)(4). DEVICE EVALUATION: G6, G6, H2, H3, H6, H10. TIGER AESTHETICS MEDICAL PERFORMED AN INVESTIGATION ON A DIFFERENT BATCH/LOT# THE REPORTED CONDITION OF FAT ON THE OUTSIDE OF THE BASKET AND INSIDE THE CHAMBER WALLS SEEMS VERY UNLIKELY TO HAPPEN. THE VACUUM NOT WORKING IS LIKELY DUE TO THE FLEXING OF THE CHAMBER WALLS AND CAUSING INTERMITTENT VACUUM. NO CAUSE COULD BE DETERMINED FOR THE REPORTED ISSUE. A VACUUM LEAK CAN DEVELOP BETWEEN THE FOAM TRAY AND SLIDING DRAWER IS A KNOWN ISSUE AND IS ADDRESSED. TIGER AESTHETICS MEDICAL WILL SUBMIT A SUPPLEMENTAL REPORT IN ACCORDANCE WITH 21 CFR SECTION 803.56 IF ADDITIONAL INFORMATION BECOMES AVAILABLE.

Additional Manufacturer Narrative · 0

TIGER AESTHETICS MEDICAL COMPLAINT #: (B)(4). PATIENT AGE DEFAULTED TO "99" AS NO PATIENT INFORMATION WAS PROVIDED. AT THIS TIME, THE SUSPECT DEVICE HAS NOT BEEN RETURNED FOR EVALUATION. TIGER AESTHETICS MEDICAL WILL SUBMIT A SUPPLEMENTAL REPORT IN ACCORDANCE WITH 21 CFR SECTION 803.56 IF ADDITIONAL INFORMATION BECOMES AVAILABLE.

Description of Event or Problem · 0

THE VIALITY UNIT SEEMED TO BE FAULTY WITH FAT GOING INTO THE SIDE WALLS, INSTEAD OF THE BASKET. SUCTION WAS NOT WORKING PROPERLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
166204 LIPOASPIRATE WASH SYSTEM FAT TRANSFER QKL TIGER AESTHETICS MEDICAL VIALITY-1400 A57455

Patients

Seq Age Sex Outcome Treatment
1